Personnel:
n
System/network administrator (recom-
mended)
Materials:
n
Network cable (EIA/TIA-568 standard),
(included in the scope of delivery of PC-
KIT-NET)
Requirements:
n
The TCP/IP module has been installed (
tional unit components” on page 18
“Assembly and con-
nection manual – PC-KIT-STICK/PC-KIT-NET”
).
n
The KIRSCH-DATANET software (version 5.0 or higher) has
been installed on the local PC or local network.
1.
Switch off the unit
Chapter 7.2 “Switching off the unit”
.
2.
Unplug the power plug.
3.
Remove the dust cover from TCP/IP output.
4.
Insert the network cable in the TCP/IP output.
5.
Connect the network cable to the network socket.
ð
The unit has been connected to the network.
6.
Insert the power plug in the socket.
7.
Switch on the unit
Chapter 7.1 “Switching on the unit”
.
1.
Insert unit key in key switch.
2.
Set unit key to “P”.
3.
Press and hold
[Max.]
and
[Min.]
simultaneously for four sec-
onds.
ð
The display shows
.
4.
Press
[Max.]
and
[Min.]
to go to level
and confirm with
Set.
5.
Press
[Min.]
to select the parameter
.
ð
The first digits of the MAC address are displayed.
Connecting the unit to the local
network
Reading out the MAC IP address on
the unit
Operation
Setting up PC-KIT-NET (optional)
26.05.2021
Refrigerator/freezer BL series
55