2.3
Extension and contraction of footrest do not move.
2.4
Heater does not work.(It does not heat up)
Is the footrest movable to maximum upper angle?
No
Is DC24V(approx..) out on the connector of main circuit
board(CN18, 19) when heater is ON.
(Check with tester.)
No
Replace the main circuit board.
Yes
Exchange heater or the cord for heater.
Is there any disconnection for each connecting part of
heater?
No
Connect the connector firmly.
Yes
Is any error occurred?
4D-970
:
“Error occurs
・
Switch the power again” is displayed on the LCD.
Check each part after confirming service No.(Error no.)
3.
In case service No.(Error No.) is displayed.
<
Reference
>
Yes