background image

10

ENGLISH

OPERATING INSTRUCTIONS

DIODE TEST

1. Insert RED test lead into VΩ jack 

5

 and BLACK test lead into COM

jack 

4

, and rotate function selector switch 

2

 to the 

 setting.

2. Touch test leads to diode. A reading of 200-700mV on display

indicates forward bias, OL indicates reverse bias. An open

device will show OL in both polarities. A shorted device will

show approximately 0mV.

BATTERY TEST

1. Insert RED test lead into VΩ jack 

5

 and BLACK test lead into

COM jack 

4

, and rotate function selector switch 

2

 to the 1.5V

or 9V battery test setting.

2. Connect BLACK lead to negative, and RED lead to positive

terminal of battery.

3. Measure voltage on display, batteries in good condition should

be within approx. 10% of rated voltage.

DATA HOLD

Press Data Hold button 

6

 to hold the measurement on the display. 

Press again to release the display and return to live measuring.   

Red lead

Black lead

Red lead

Black lead

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Summary of Contents for 69149P

Page 1: ...L ENGLISH MM300 600V 10A 2M DATA HOLD AUDIBLE CONTINUITY BATTERY TEST DIODE TEST Manual Ranging Digital Multimeter Digital Multimeter www GlobalTestSupply com Find Quality Products Online at sales Glo...

Page 2: ...r one year Standards Conforms to UL STD 61010 1 61010 2 030 61010 2 033 Certified to CSA STD C22 2 61010 1 61010 2 030 61010 2 033 IEC EN 61010 1 61010 2 030 61010 2 033 61326 1 Pollution degree 2 Acc...

Page 3: ...ance 200 0 0 1 1 0 5 digits 2000 1 20 00k 0 01k 200 0k 0 1k 2000k 1k 1 5 5 digits Battery Test 9V 10mV 1 0 2 digits 1 5V 10mV 1 0 2 digits Diode Test Approx 1mA open circuit voltage 2 0V DC Continuity...

Page 4: ...e a shock hazard To avoid false readings that can lead to electrical shock replace batteries when a low battery indicator appears Do not attempt to measure resistance or continuity on a live circuit A...

Page 5: ...count LCD display 2 Function selector switch 3 10A jack 4 COM jack 5 V jack 6 HOLD Data Hold button NOTE There are no user serviceable parts inside meter www GlobalTestSupply com Find Quality Products...

Page 6: ...T LOCATIONS Ensure the test lead shield is pressed firmly in place Failure to use the CAT III CAT IV shield increases arc flash risk TESTING IN CAT II MEASUREMENT LOCATIONS CAT III CAT IV shields may...

Page 7: ...ching the test leads together to close the circuit the meter will measure zero volts NOTE Do not attempt to measure more than 600V or 200mA OPERATING INSTRUCTIONS AC VOLTAGE LESS THAN 600V 1 Insert RE...

Page 8: ...t Remove power from circuit open circuit at measurement point connect meter in series in the circuit using the test leads and apply power to circuit The meter will auto range to display the measuremen...

Page 9: ...r is under test the display will indicate O L This is normal DO NOT attempt to measure resistance on a live circuit CONTINUITY 1 Insert RED test lead into V jack 5 and BLACK test lead into COM jack 4...

Page 10: ...Insert RED test lead into V jack 5 and BLACK test lead into COM jack 4 and rotate function selector switch 2 to the 1 5V or 9V battery test setting 2 Connect BLACK lead to negative and RED lead to pos...

Page 11: ...00mA 600V fast blow Klein Cat No 69031 10A jack 3 10A 600V fast blow Klein Cat No 69032 3 Replace back housing and fasten securely with screws To avoid risk of electric shock disconnect leads from any...

Page 12: ...d of storage in extreme conditions exceeding the limits mentioned in the General Specifications section allow the meter to return to normal operating conditions before using DISPOSAL RECYCLE Do not pl...

Reviews: