Troubleshooting 

1.8.1

 

KCS-ID messages 

Amplifier Headroom exceeded 

The  KA3’s  amplifier  can  provide  approx.  19…22  V  peak  (depending  connected  DC  resistance).  The 
error message pops up if more voltage is required to reach a certain limit or the excitation voltage is 
set manually to a level exceeding the amplifier’s capability. Note that the multi-tone stimulus has a 
crest factor of approx. 3. Hence, the maximum possible voltage is 6…7 V RMS when the internal AMP 
card is used. 

You can use an external amplifier that can provide more voltage to drive the transducer to its limits 
or to reach a higher excitation voltage. 

If 6…7 V RMS are sufficient because the target amplifier used in the KCS online mode cannot deliver 
more voltage anyway, you can set  the  level manually to approx. 6 V  RMS  in the  KCS-ID’s property 
page’s category 

Stimulus. 

 

High voice coil position change detected 

This message indicates that between LINEAR MODE and NONLINEAR MODE either 

 

the  voice  coil  was  shifted  significantly  due  to  very  asymmetric  nonlinear  behavior  of  the 
transducer or 

 

the distance between laser sensor and transducer membrane position has changed 

Note that the rest position of the NONLINEAR MODE is used as reference rest position used in the 
KCS on-line mode. The voice  coil stabilization feature  will shift  the voice coil to this reference  rest 
position. 
 

Check correct Speaker System Type setting 

KCS-ID  performs  a  small  signal  speaker  parameter  identification  in  the  LINEAR MODE.  If  it  detects 
that e.g. the number of distinct impedance magnitude peaks does not  correspond to the specified 
Speaker  System  Type  (e.g.  sealed  or  vented  enclosure),  this  message  appears.  Note  that  in  rare 
cases, the warning might trigger even if the correct speaker type is set. In this case you do not have 
to do anything. 

1.8.2

 

KCS Server messages 

The  KCS  Server  website  might  display  warnings  and  information  after  a  KCS-ID  file  has  been 
uploaded. See explanations of selected info and error messages below. 

Info/Warning Messages 

Low  both-sided  Bl(x)  decay.  Detection  of  the  voice  coil  rest  position  and  stabilization  might  not 
work optimally.

 

Summary of Contents for KCS-ID

Page 1: ...R D SYSTEM USER S GUIDE KCS ID Manual K l i p p e l G m b H July 23 2021 Copyright 2000 2021 Klippel GmbH Mendelssohnallee 30 01309 Dresden Germany www klippel de...

Page 2: ......

Page 3: ...urements 3 1 4 1 Basic Measurement Setup 3 1 4 2 Additional Options 4 1 4 3 Limit Parameters 4 1 4 4 Running and Interpreting Measurements 5 1 5 Creating Initial Data 7 1 6 Xmax and Xtarget 7 1 6 1 Wh...

Page 4: ......

Page 5: ...is initial data set is stored in a KCS Monitor operation and can be downloaded from the server The KCS Monitor operation can connect to any supported KCS hardware platform and a KCS control session ca...

Page 6: ...prises all relevant information required by KCS ID Make sure that IEPE phantom power is switched on using the buttons next to the microphone input on the KA3 s front panel If a transducer with multipl...

Page 7: ...like used in the KCS on line mode 2 Select Limit Parameters It is recommended to never switch off the Impulsive Distortion limit which indicates excessive impulsive distortion Rub Buzz If the target...

Page 8: ...have access to the transducer s membrane and cannot be used Note that a microphone calibration is not required for this measurement Advanced Save Intermediate Results saves the results of each measure...

Page 9: ...change the default impulsive distortion limits Note The IDR is very susceptible for impulsive distortion e g voice coil rubbing and not so much for air flow noises which exhibit less impulsive charact...

Page 10: ...t has successfully finished the measurement status overview in the State table displays a PASS after all measurement modes and an 8 character identification number Initial KCS data can now be created...

Page 11: ...tor KCS3 Before evaluating KCS it is highly recommended to see Tutorial 4 Alignment Equalization and 5 Protection Limiter of KCS3 which describe steps how to achieve best audio performance Note The ha...

Page 12: ...which can cause an asymmetric peak excursion This means that for instance the positive peak displacement is higher than the absolute peak value on the negative side see curves below If the voice coil...

Page 13: ...multiple voice coils each driven by a separate amplifier channel Note KCS only supports multi coil configurations if all coils show identical small and large signal properties Please contact Klippel...

Page 14: ...INEAR MODE either the voice coil was shifted significantly due to very asymmetric nonlinear behavior of the transducer or the distance between laser sensor and transducer membrane position has changed...

Page 15: ...using a DC voltage Stabilization is not possible Note that KCS will work anyway but higher safety margins for the maximum allowed excursion are required leading to less acoustical output at low freque...

Page 16: ...educed The gain of the specified target amplifier is not sufficient to drive the speaker to the Xmax determined in the KCS ID Xmax was reduced Error Messages High Linear Identification error High Nonl...

Page 17: ...Setup Routing 2 How to achieve a higher Xmax 8 I Info Warning Messages 10 Introduction 1 K KCS Server messages 10 KCS ID Manual 1 KCS ID messages 10 L Limit Parameters 4 P Performing Measurements 3 R...

Reviews: