21
Installation and initial startup
AZURA® Detector UVD 2.1L instructions, V6831
4.9.2 Mobile Control: Setting a dynamic IP address via
device name
The device is switched on.
Mobile Control is installed and running.
The connection between Mobile Control and the device has been
established.
Process
1.
In Mobile Control, click <Settings>.
2.
On the <General> tab, choose the device.
3.
In <Network Settings>, choose the setting <DHCP>
1
.
4.
Click in the top right corner.
5.
Restart the device (recommended).
1
IP address mode
Abb. 14:
Network settings for dynamic IP address
The device is now accessible via a dynamic IP address.
4.9.3 Mobile Control: Setting a dynamic IP address via
device serial number
The device is switched on.
Mobile Control is installed and running.
The connection between Mobile Control and the device has been
established.
Process
1.
In Mobile Control, click <Settings>.
2.
In <Network Settings>, click <Reset>. The windows <Reset commu-
nication settings> opens.
3.
Enter the serial number of the device into the text field.
4.
Click <OK>. The device is now reset to default settings.
5.
Restart the device (recommended).
The device is now accessible via a dynamic IP address.
Prerequisites
Procedure
1
Result
Prerequisites
Procedure
Result