28
AZURA® Detector UVD 2.1L Instructions, V6831
5. Operation
Note:
Before the detector is ready to use for measurements, a flow cell
must be installed. You can change the optical path length of a preparative
flow cell before assembling the flow cell. The optimal path length de-
pends on type and quantity of the sample.
5.1 Checklist before operation
Device is in a location which suits the ambient conditions and space
requirements specified in section „Ambient conditions“ on page 11.
The power connection of the detector is plugged in
(see „4.3 Power supply“, p. 12).
If the detector is part of a HPLC system, the following must be observed:
The network connection to the router is established
(see „4.8.4 Configuring the router“, p. 18).
A chromatography software has been installed by KNAUER or a com-
pany authorized by KNAUER.
Capillaries from the column to the UV detector and capillaries from the
detector to the waste bottle are securely attached.
5.2 Software control
A device can be operated in several ways:
with chromatography software
with Mobile Control
Note:
It is not possible to use two operating methods simultaneously. If
the device is connected to the software, it cannot be controlled via Mobi-
le Control. The device status can however be monitored.
5.2.1 Chromatography software
To control the device with chromatography software, it must be connec-
ted to the computer through the LAN interface.
Devices can be controlled with e. g. OpenLAB® EZChrom edition version
A.04.05 or higher, ChromGate version 3.3.2 or higher and ClarityChrom®
version 3.0.7 or higher, Chromeleon® 7.2 and PurityChrom® version
5.07.039 or higher.
You will find a detailed description on the chromatography software in
the software instructions.
5.2.2 Mobile Control
The Mobile Control is a device control software which can be installed on
your computer or tablet. To control the device using the Mobile Control,
connect the computer or tablet to a wireless LAN router. Data transfer
between device and Mobile Control is actualized through wireless LAN.
You find a detailed description on the Mobile Control in its accompany-
ing instructions.