Maintenance and care
AZURA® Detector UVD 2.1L Instructions, V6831
Mount the flow cell onto the detector.
8.5 Replacing the fiber optics
UV light will cause the fiber optics to become blind with time (solariza-
tion), making them no longer suitable for use.
Note:
Observe the following regarding the use of fiber optics:
Do not touch the ends of the fiber optics with your fingers, as this
could falsify the measurement.
Handle the fiber optics with care and avoid impacts.
Move the fiber optics carefully without using pressure or bending it.
Detector is switched off.
Cap fittings for fiber optics
Steps
Figure
1.
To detach from the flow
cell, unscrew the fittings
of the fiber optics manu-
ally
1
.
Fig. 24
Fiber optics fittings on flow cell
2.
Seal the fiber optic con-
nectors with caps
2
.
Fig. 25
Cap for optic connector
3.
Manually, unscrew the
fiber optics
3
from the
detector.
4.
To mount new fiber
optics onto the flow cell,
repeat step 1-3 in re-
verse order.
Fig. 26
Fiber optics on the detector
The fiber optics are replaced.
Next steps
Prerequisite
Tools
Result