Liquid pulsation damper
Installation and connection
KNF Flodos BA_FPD_EN_04_069182
Translation of Original Operating and Installation Instructions
11
Be aware of the following before commissioning the liquid pulsation
damper:
Follow the safety regulations specific to the media used.
1.
The materials of the liquid pulsation damper, the hose con-
2.
nections and the hoses themselves must be sufficiently
chemically resistant to the liquid being transferred (see Chap-
ter 4.1, Tab. 3).
Check the correct and tight connection of hoses and any
3.
transition joints (hose connectors / hose, etc.).
We recommend securing transition joints between hose and
hose connectors with hose clamps or other clamping elements.
If you are working in an area under pressure, this is imperative
to prevent the hoses from coming off the hose connectors.
In the case of special models, the internal nozzle cross-section
(Ø) specified in the system parameters is noted on the type
plate, e.g.
0 D1.2.
The liquid pulsation damper can be fastened from above and
below using the four holes.
It can be mounted in any orientation.
Hydraulic
6.2.
Remove the protective caps from the connections.
Mount the hose connectors (observe the recommended hose
inner diameter of the liquid pulsation damper).
The hydraulic connections (supply and output sides) are de-
fined using directional arrows on the upper section.
Push the hoses on the supply and output sides over the hose
connectors and secure.
Check that the system is leak-tight.
The liquid pulsation damper, hoses and other components
must be rated for the maximum permissible operating pres-
sure.
General information
Mechanical