Diaphragm Vacuum Pumps N 920__.18 and N 920 KT.29.18G
Operation
Translation of original Operating Instructions, English, KNF 121416-121419 07/18
15
7.5. Operating the pump via VC 900
(only N 920 KT.29.18G)
If the interface cable (accessories) is connected, the pump can be
operat ed via t he vacuum controller VC 900. For f urther inf or-
mation on operating see operating instructions VC 900.
7.6. Gas ballast
WARNING
Personal injury caused by poisoning or explosion
and damage to the pump
Make sure that no reactive or explosive mixtures
will be produced when the gas ballast valve is
open.
When transferring vaporous media, opening the gas ballast
valve can minimize the formation of condensation in the pump
heads.
Opening the gas ballast valve deteriorates ultimate vacuum
performance.
The gas ballast valve is opened and closed via the gas ballast
switch (see Fig. 5).
Please contact KNF Service if you require an inert connection for
the gas ballast.
Fig. 5: Gas ballast switch