www.grupotemper.com 13 14 www.grupotemper.com
KMD-20AFC-01
PC INTERFACE AUTO-RANGE DIGITAL MULTIMETER
KMD-20AFC-01
PC INTERFACE AUTO-RANGE DIGITAL MULTIMETER
MEASURING DUTY CYCLE
1. Connect the black test lead to the (-) jack COM and red lead to
the (+) jack (Hz).
2. Set the function switch to "Hz/% duty”.
3. Push the “Hz/%” button changing the function to % duty cycle.
4. Connect the test leads to the circuit to be measured. The range
will change that will display the measured duty cycle with the best
resolution.
MEASURING TEMPERATURE
1. Set the function switch to "TEMP
℃
/
℉
”.
2. Connect sensor positive to the “uA mA” terminal and sensor
negative to “COM” terminal
3.Connect K-Type test line to the sensor (positive to positive,
negative to negative)
4. Test line to the test point being measured
5. Read the temperature value (
℃
or
℉
) on LCD
RS-232 PC INTERFACE MEASUREMENTS
1. Connect the RS-232 wires to the meter’s “RS-232” interface and
PC “COM” port.
2. Presses the meter’s RS232 button and the indicator “RS232” will
be appear in the display.
3. PC will display the meter’s measuring model.
Note:
First of all please read the “README” file in the
attached soft disk and then comply with it set the SET UP
Program from soft disk up to your PC.
BATTERY AND FUSE REPLACEMENT
If the sign “BAT” appears on the LCD display, it indicates that the
battery should be replaced. Remove screws on the back cover and
open the case. Replace the exhausted battery with new battery.
( Standard 9V battery NEDA 1604, IEC 6F22)
Fuse rarely need replacement and blow almost always as a result of
the operator’s error. Open the case as and replace the blown fuse
with ratings specified.
WARNING:
Before attempting to open the case, be sure that test leads have
been disconnected from measurement circuit to avoid electric shock
hazard.
Replace fuse only with specified ratings:
Fuse: F1 500mA / 250V fast blow, F2 10A / 600V fast blow.