DVZ
page 8
DVZ K04/0810
7.8 DVZ-...Ex4R, DVZ-...Gx4R
Cable connection
Wire number
DVZ-...E14R
Counter electronics
DVZ-...G14R
Dosing electronics
1 +24
V
DC
+24
V
DC
2 GND
GND
3
4-20 mA
4-20 mA
4
GND
GND
5
n. c.
control 2*
6
reset part quantity
control 1*
7 relay
S1
normally open
relay S1
normally open
8
9
relay S2
normally open
relay S2
normally open
10
*Start-dosing: Control 1 connect with GND
Stop-dosing: Control 2 connect with GND
Reset-dosing: Control 1 and Control 2 at the same time with GND
Do not connect any external voltage to the control units!
Plug connection
-E34 R
+ Vs
Reset TM
0 (4) - 20 mA
GND
1
30 V / 2 A
AC/DC
3
4
5
1
3
4
5
6
S 2
8
d.c. *)
Output
d.c.*) Don't connect!