User manual - DATA RECORDER and cooperating software
”.
”
Close delay
” - Output switch from open to closed state will be delayed until this time elapse.
”
Open delay
” - Output switch from closed to open state will be delayed until this time elapse.
If condition to switch state of the output will repeat occurring more often than half of
the delay and mean duty cycle (
D
) will be greater than 50% the output will
eventually switch. See chapter ”
” for more details.
”
Closed hold time
” - a relay active state maintain time after the triggering signal disappears,
”
Open hold time
” - a relay inactive state maintain time after the triggering signal disappears,
• All time type parameters which describe outputs, can be set in range 0 ÷ 9999
with resolution of 1 sec.
• Because of computing precision, the time is shorter about 2.5% than the entered
value. This means if time is set to 100.0s the real time will be shorter about 2.5s.
” menu.
7.4.8.
”Date & time settings” menu
Menu allows to change date and time of internal clock (RTC):
”Time”
- current time according to the unit’s real time clock,
”Date”
- current date according to the unit’s real time clock,
”Apply”
- this option enables the internal real time clock to be set according to values of the
“
Time
” and “
Date
” parameters.
After entering the menu the current indications of the internal clock are displayed. The
changing of parameters (
“Time”
or
“Date”
) stops the indications updating. Changed
parameters must be stored using the
“Apply”
option.