DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
103 of 326
Error 12-080: Software Update Error
Messages
Description
The LOCAL PANEL cover screen was notified that the software update was not successfully
completed. The error is logged, and it cannot be cleared until the system is “re-ghosted”.
Result of Error
The system is unusable because the failed software update could have left the system in any
state. It is unknown if diagnostic quality images can be generated. Images will not be printed.
User Action
None
FE Action
None, this is a log notification only.
LOCAL PANEL
LOG
DICOM
SEVERITY
None
(No alarm)
Software Update Failure
NORMAL
1-Alert