DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
171 of 326
Error 21-932, 22-932, or 23-932: SUPPLY CONTROLLER Communications
Retry
Messages
Description
Note
UPPER DRAWER (U) = 21. MIDDLE DRAWER (M) = 22, LOWER DRAWER (L) = 23.
This error occurs when the MCS tries to communicate with the MICROCONTROLLER but
receives no response. The error is cleared when power is cycled on the IMAGER.
Note
Error 2x-931 is generated when this problem is detected after 5 successive communications
attempts.
Result of Error
None
User Action
None
FE Action
None, this is a log notification only.
LOCAL PANEL
LOG
DICOM
SEVERITY
None
(No alarm)
MCS failed to communicate
with xxxxx Supply
NORMAL
3-Notify