DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
180 of 326
Error 26-544: Jam at DENSITOMETER
Messages
Description
This error occurs if the trailing edge is not detected at the DENSITOMETER FILM DETECT
SENSOR (S18) after the film exits the DENSITOMETER. The error is cleared when the
IMAGER DOORS are opened then closed. Open is indicated by either I3 (PROCESSOR and
vertical DOOR) or I4 (SORTER) INTERLOCKS open. Close is indicated by both I3 and I4
INTERLOCKS closed. When this error occurs, the PROCESSOR stops rotating and the
HEATER is turned off. All films in the PRINTER will be marked bad and reprinted
automatically. This includes films in the lower part of the PRINTER that will also be cleared
automatically when the IMAGER is locked.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Remove any films in the SORTER.
2
Use the FILM FEED KNOB in area 5 to advance the film into to SORTER.
3
Lift the SORTER section and remove any films.
4
If the error occurs again, call Service.
LOCAL PANEL
LOG
DICOM
SEVERITY
Film Jam in Area 5
(Alarm sounds)
Film Jam at Densitometer:
Film at Densitometer sensor
(S18) failed to detect film
trailing edge
FILM JAM
1-Alert