DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
188 of 326
Error 27-601: Invalid POWER MONITOR Offset
Messages
Description
This error occurs when the POWER MONITOR offset measured by the OPTICS calibration
and diagnostics is outside the range. The error clears when a calibration is asked for or
power is cycled.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
DICOM
SEVERITY
Optics: Calibration Failed
(Alarm sounds)
Optics Calibration Failed.
Power Monitor Offset outside
valid range
ELEC DOWN
0-Fatal