DIAGNOSTICS
Using the Diagnostics
30JUL07
8F2919
Page
19 of 326
OPTICS
Starting the Tests
1
Select
Diagnostics>MCS>Optics
.
OPTICS Calibration
This calibration of the OPTICS is the same that occurs before a calibration print or when the
IMAGER is energized. The calibration:
•
Sets the SPINNER MOTOR to full speed and checks that it is at full speed.
•
Checks laser offset and gain, and adjusts if necessary.
•
Initializes the POWER MONITOR by measuring offset and “range”.
•
Does calibration of the ATTENUATOR by measuring it at the full “range”.
•
Does calibration of the LASER by measuring the “range” and maximum power.
1
To do a calibration of the OPTICS, select “Calibrate Optics” from the “Optics Calibration
and Diagnostics” menu.
2
Click [Run].
When the calibration completes, the SCREEN will display “Status” and a “Summary” of
the Calibration.
Displaying Current Calibration Data
This function displays the data from the last calibration.
1
Select “Display Current Calibration Data.”
2
Click [Run].
“Exporting” Calibration Data
This procedure allows you to send calibration data to your LAPTOP COMPUTER.
1
Select “Export Calibration Data.”
2
To format the data, click [Run].
3
To save the data to a folder on your LAPTOP COMPUTER: