DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
202 of 326
Error 27-608: Invalid LASER Dynamic Range
Messages
Description
This occurs when the laser dynamic range is adjusted and the value is within 10% of the
adjustment range.
Result of Error
None
User Action
None
FE Action
None, this is a log notification only.
LOCAL PANEL
LOG
DICOM
SEVERITY
None
(No alarm)
Optics Calibration detected
laser dynamic range
approaching predefined limits
NORMAL
3-Notify