DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
208 of 326
Error 27-647: START OF PAGE SENSOR Failure
Messages
Description
This error occurs when the START OF PAGE SENSOR (S13) does not detect the normal
LASER on or off state during diagnostics. The error is cleared when power is cycled on the
IMAGER.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
DICOM
SEVERITY
Optics: Internal Hardware
Failure
(Alarm sounds)
Start of Page Sensor (S13)
failed diagnostics
ELEC DOWN
0-Fatal