DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
222 of 326
Error 28-501: PROCESSOR RTD Short Circuit
Messages
Description
This error occurs when the temperature calculated by the PROCESSOR
MICROPROCESSOR is almost zero. The temperature is calculated from the voltage
measured by the RTD. A possible cause is a short in the RTD circuit. The RTD number is
included in the error log message. The error is cleared when power is cycled on the
IMAGER.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
DICOM
SEVERITY
Processor: Internal Hardware
Failure
(Alarm sounds)
Processor calculates
temperature near zero.
Probably short in RTD circuit
PROC DOWN
0-Fatal