DIAGNOSTICS
Using the Diagnostics
30JUL07
8F2919
Page
24 of 326
1
To check operation of the SORTER SENSOR, select
Diagnostics>MCS>Sorter
.
2
To do calibration of the SORTER SENSOR:
a
Set the SERVICE SWITCH in the “Service” mode.
b
In the top right corner of the screen, click [Run].
The calibration will run and indicate OK. The results will display on the “Monitor
Sensor Charts”. You must use the “charts” to determine if a failure occurred. For
example, if the EMITTER does not cause enough light at the DETECTOR, the
“Sensor Reading” will be -1 and the “Sensor LED Current” will be 255.
c
To end monitoring, on the lower part of the screen, click [Stop].
3
To display the output of the SORTER SENSOR with the results of the last calibration:
a
On the “Monitor Sensor” area of the screen, click [Run].
b
To stop monitoring, click [Stop].
4
After completing the diagnostic, set the SERVICE SWITCH in the user position and
restart the IMAGER.