DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
241 of 326
Error 29-927: DENSITOMETER Fiducial Failed Reference Diagnostics
Messages
Description
During a calibration print and during DENSITOMETER diagnostics, a reading from the
FIDUCIAL SENSOR (S17) is taken with the light source turned on. This error occurs when
the reading is out of range. The error is cleared when power is cycled on the IMAGER.
Note
Error 29-947 is generated when this reading is reaching the range limit.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
DICOM
SEVERITY
Densitometer: Internal
Diagnostic Failure
(Alarm sounds)
Densitometer fiducial sensor
reference reading out of
range with light source on
ELEC DOWN
0-Fatal