DIAGNOSTICS
Using the Diagnostics
30JUL07
8F2919
Page
26 of 326
Film Calibration Data
1
To obtain a “DLogE Curve” from the last calibration data for film in any CARTRIDGE in
the IMAGER, select
Hardware Data>Film Calibration
.
2
Select the SUPPLY DRAWER you want to check.
3
Click:
•
[Get Records]
•
[Plot]
A “DLogE Curve” of the characteristics of the film in the selected DRAWER will display.
The “curve” is “plotted” from data stored from the last calibration for this film.
4
To make a copy of the “DLogE Curve” to your LAPTOP, click [Export Data].
5
Copy the file to a folder on the HARD DRIVE of your LAPTOP COMPUTER.