DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
263 of 326
Error 32-301: FILM REGISTRATION Centering Failure
Messages
Description
This error occurs when the REGISTRATION CENTERING HOME SENSOR (S10) does not
detect successful position of the CENTERING ARMS by the CENTERING MOTOR (M7). This
error can occur at multiple points in the FILM REGISTRATION. The film will not be exposed,
but will be moved out of the IMAGER. The error is cleared when power is cycled on the
IMAGER.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
DICOM
SEVERITY
Film Registration Failure
(Alarm sounds)
Film Registration Centering
Failure: Centering Home
sensor S10 indicates
Centering motor M7 failure
FILM TRANS
ERR
1-Alert