Error 32-304: Internal Temperature/Humidity SENSOR Failure
Messages
Description
This error occurs if the TEMPERATURE/RH SENSOR (S25) failed diagnostics.
Result of Error
None
User Action
None
FE Action
1
De-energize then energize the LASER IMAGER.
2
Check and seat all CONNECTORS on the FRB.
3
Use the SERVICE TOOL to check the TEMPERATURE/RH SENSOR.
4
Install a new FRB. See the ADJUSTMENTS AND REPLACEMENTS,
REGISTRATION BOARD (FRB) - FILM REGISTRATION AY
.
LOCAL PANEL
LOG
DICOM
SEVERITY
None
(No alarm)
Internal Temperature/
Humidity Sensor Failure
NORMAL
3-Notify