background image

DIAGNOSTICS

Status Messages and Error Codes

30JUL07

8F2919 

Page 

273 of 326

Error 33-311: EXPOSURE TRANSPORT ENTRANCE FEED ROLLER Failure

Messages

Description

This error occurs when the EXPOSURE ENTRANCE CLAMP SENSOR (S14) does not detect 
successful position of the lower ROLLERS by the EXPOSURE ENTRANCE CLAMP MOTOR 
(M10). This error can occur at multiple points in the EXPOSURE TRANSPORT process. The 
error is cleared when power is cycled on the IMAGER.

Result of Error

The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new 
films will not be initiated. Films downstream of the subsystem will be completed. All other 
films will be reprinted.

User Action

1

Restart the LASER IMAGER from the LOCAL PANEL.

2

If the error occurs again, call Service.

FE Action

Do the actions below in sequence. If an action provides a resolution, do not continue with the 
other actions.

1

De-energize then energize the LASER IMAGER.

LOCAL PANEL

LOG

DICOM

SEVERITY

Exposure Transport: Entrance 
Failure
(Alarm sounds)

Exposure Transport Entrance 
Feed Roller Failure: Entrance 
Roller Position sensor S14 
indicates Exposure Transport 
Entrance Motor M10 failure

FILM TRANS 
ERR

1-Alert 

Summary of Contents for DryView 6800

Page 1: ...ream Health Inc 2007 Diagnostics Production HealthGroup ExternalAndInternal Publication No 8F2919 30JUL07 DIAGNOSTICS for the Kodak DryView 6800 LASER IMAGER Service Code 1649 Important Qualified personnel must repair this equipment H210_0800GC ...

Page 2: ...egligence or other fault This equipment includes parts and assemblies sensitive to damage from electrostatic discharge Use caution to prevent damage during all service procedures Description Page Table of Contents Using the Diagnostics 8 Using the SERVICE SWITCH 8 SERVICE TOOL 9 Functions 9 Connecting Disconnecting the LAPTOP COMPUTER to from the IMAGER 10 Logging On to the SERVICE TOOL 11 Logging...

Page 3: ...TEM Error 54 Error 04 100 Limit on Number of Associations Exceeded 55 Error 04 101 AE Title does not Correspond to a Recognized PRINTER 56 Error 04 110 Invalid Association Request 57 Error 04 200 DISK Full 58 Error 04 201 Keep Alive Message Failure 59 Error 04 202 Connection Is Lost 60 Error 04 300 PLUT is Missing 61 Error 04 301 Missing Necessary Attribute 62 Error 04 302 Invalid Attribute 63 Err...

Page 4: ... 100 Error 12 075 LOCAL PANEL Error 101 Error 12 080 Software Update Error 102 Error 12 085 LOCAL PANEL Flash Error 103 Error 13 001 SERVICE TOOL Internal Error 104 Error 13 002 SERVICE TOOL Database Error 105 Error 20 004 USB Did Not initialize 106 Error 20 154 MCS HARD DRIVE Failure 107 Error 20 156 Incompatible Software Versions Installed 108 Error 20 209 LASER IMAGER Opened During Self Test 11...

Page 5: ... Error 21 931 22 931 or 23 931 SUPPLY CONTROLLER Communications Failure 168 Error 21 932 22 932 or 23 932 SUPPLY CONTROLLER Communications Retry 170 Error 26 321 Jam Exiting ACCUMULATOR 171 Error 26 325 Jam Entering EXPOSURE TRANSPORT 173 Error 26 326 Jam Exiting EXPOSURE TRANSPORT 175 Error 26 543 Jam Before the DENSITOMETER 177 Error 26 544 Jam at DENSITOMETER 179 Error 27 121 ATTENUATOR MOTOR H...

Page 6: ...or 29 947 DENSITOMETER Fiducial Reference Diagnostics Warning 246 Error 30 154 SORTER NVRAM Failure 247 Error 30 931 SORTER Communications Failure 249 Error 30 932 SORTER Communications Retry 251 Error 30 935 SORTER DIVERTER 1 Failure 252 Error 30 936 SORTER DIVERTER 2 Failure 254 Error 30 937 SORTER DIVERTER 3 Failure 256 Error 30 938 SORTER DIVERTER 4 Failure 258 Error 32 154 FILM REGISTRATION N...

Page 7: ...2 Processor Transport Communications Retry 302 Troubleshooting 303 System Problems 303 POWER SUPPLY Problems 304 Functions of the MICRO BOARD LEDs Test Points and SWITCHES 305 Happy LEDs on the MICRO BOARDS 305 POWER DISTRIBUTION BOARD PDB 306 FILM SUPPLY BOARD FSB 307 FILM REGISTRATION BOARD FRB 308 EXPOSURE TRANSPORT BOARD ETB 309 PROCESSOR CONTROL BOARD PCB 312 SORTER CONTROL BOARD SCB 313 Usin...

Page 8: ...the SERVICE SWITCH The SERVICE SWITCH is located inside the MAIN DOOR behind the LATCH MOTOR 1 Open the MAIN DOOR see the INSTALLATION INSTRUCTIONS Using the Manual LATCH RELEASE 2 Loosen the top SCREW of the BRACKET 3 Remove the bottom SCREW H210_0582AC SERVICE POSITION Top SCREW BRACKET Bottom SCREW ...

Page 9: ...vides Logs Provides access to logs of events that occur in the IMAGER Monitoring and Control Provides information about the DRE print counts SENSORS and PROCESSOR temperatures System Information Provides information about the IMAGER the service user and software versions Upgrade Provides the ability to upgrade the DRE software and the MICROCONTROLLER applications of the MCS Utilities Provides the ...

Page 10: ...ELECTRONICS FILTER c Connect the LAPTOP COMPUTER to the RJ45 CONNECTOR on the DRE with a CROSSOVER CABLE TL 5568 Note If you are printing films while the LAPTOP COMPUTER is connected to the IMAGER you can ensure that the IMAGER is light tight by routing the CROSSOVER CABLE as shown in the picture below then closing the AIR INTAKE DOOR AIR INTAKE DOOR ELECTRONICS FILTER 2 SCREWS RJ45 CONNECTOR AIR ...

Page 11: ... Using the Diagnostics 30JUL07 8F2919 Page 11 of 326 2 To disconnect a Disconnect the CROSSOVER CABLE from the IMAGER b Install ELECTRONICS FILTER 3 NUTS c Close the AIR INTAKE DOOR CROSSOVER CABLE AIR INTAKE DOOR ...

Page 12: ...connecting the LAPTOP COMPUTER to from the IMAGER 2 Energize the LAPTOP COMPUTER 3 Select Start Programs Kodak SecureLink 4 Type your password 5 Click OK 6 Type the IP address of the SERVICE PORT 192 168 000 001 7 Click Connect 8 Select Start Programs Kodak Service WebLink The main menu of the SERVICE TOOL displays Logging Off From the SERVICE TOOL 1 In the upper right corner of the SERVICE TOOL c...

Page 13: ...E MANAGEMENT SERVICES 1 Connect the LAPTOP COMPUTER to the IMAGER see Connecting Disconnecting the LAPTOP COMPUTER to from the IMAGER 2 Log on to the SERVICE TOOL see Logging On to the SERVICE TOOL 3 Select Configuration System Remote Management 4 Click Modify 5 Under the Remote Services Activation section use the drop down menu to select Remote Monitoring and Remote Access 6 To ...

Page 14: ...ght values for the LAMPS 8 times and calculates an average The average value must be 4096 400 De energizes the DENSITY LAMP then reads the A to D counts for reference brightness 8 times and averages the values If the average A to D count is not between 750 000 and 850 000 software adjusts the count De energizes the DENSITY LAMP and energizes the FIDUCIAL LAMP Reads the Reference brightness value 8...

Page 15: ... Diagnostics 2 Set the SERVICE SWITCH in the service position 3 From the Select Operation drop down menu select the MOTOR you want to test 4 Click Run When the test is completed either Pass or Fail will be indicated and test status will display 5 Click the bottom line of the screen to display test results data 6 After completing the test set the SERVICE SWITCH to the user position and restart the ...

Page 16: ...t Memory This diagnostic checks image memory by writing to the memory and then reading it back Reset This function resets the selected MICROCONTROLLER to the original condition Note If you select Memory from the Cycle drop down menu select if the test should be run for One Cycle or Continuous 3 From the Select Operation drop down menu select the MICROCONTROLLER you want to test Note If you select ...

Page 17: ...ller Close This test checks the FEED ROLLER OPEN CLOSE MOTOR and SENSOR Feed Roller Open This test checks the FEED ROLLER OPEN CLOSE MOTOR and SENSOR Feed Motor This test checks the PICKUP FEED ROLLER DRIVE MOTOR Vacuum Pump and Valve This test checks the VACUUM PUMP and VALVE RF Tag This test causes the MCS to send a command to read data from the RF TAG The only data returned is the Serial Number...

Page 18: ...ted the test will run until Stop is selected Continuous should only be used for test and film transport to a sorter bin 6 Click Run After the test completes Pass or Fail will be indicated and status will display 7 Click on the bottom line of the screen to display test results 8 After completing the test set the SERVICE SWITCH to the user position and restart the IMAGER ...

Page 19: ...OR by measuring it at the full range Does calibration of the LASER by measuring the range and maximum power 1 To do a calibration of the OPTICS select Calibrate Optics from the Optics Calibration and Diagnostics menu 2 Click Run When the calibration completes the SCREEN will display Status and a Summary of the Calibration Displaying Current Calibration Data This function displays the data from the...

Page 20: ...RIVE of your LAPTOP COMPUTER OPTICS MOTORS Tests This diagnostic includes tests of the ATTENUATOR MOTOR and TRANSLATION MOTOR Each test commands the MOTOR to move a preset distance and then checks results 1 Set the SERVICE SWITCH in the service position 2 Select Test Optics Motors 3 From the Assembly Name drop down menu select the MOTOR you want to test either attenuator or translation 4 Click Run...

Page 21: ...ostics 30JUL07 8F2919 Page 21 of 326 SENSORS 1 Select Monitoring and Control Sensors Monitor 2 Select the check box of the SENSOR to be monitored Note Up to 3 SENSORS can be monitored 3 Click Start 4 To stop monitoring click Stop ...

Page 22: ...gnostics DRE Test Print b Select an image to print from the drop down menu c Click Run d Observe the Monitor Sensors screen to see changes in condition of the SENSORS 7 To see a plot of the operation of up to 3 selected SENSORS on the Graph Sensors screen a Start a test print b When the print is running return to the Monitor Sensors screen and select the SENSORS you want to monitor c Click the Gra...

Page 23: ...RRENT Decreases the threshold to 350 mV below the set DETECTOR maximum MONITOR SENSOR This test reads the output of the SORTER SENSOR and displays this reading with the values recorded in the last calibration for Threshold Dark Value Light Value and LED EMITTER Current All values are in counts from 0 255 which indicate a range of 0 5 V from the SENSOR 0 is darkest and 255 is brightest For example ...

Page 24: ...tor Sensor Charts You must use the charts to determine if a failure occurred For example if the EMITTER does not cause enough light at the DETECTOR the Sensor Reading will be 1 and the Sensor LED Current will be 255 c To end monitoring on the lower part of the screen click Stop 3 To display the output of the SORTER SENSOR with the results of the last calibration a On the Monitor Sensor area of the...

Page 25: ...re Data Film Cartridge Data 2 From the Select Tray drop down menu select the DRAWER from which you want film information Upper Middle or Lower Data from the RF TAG describing the film in the CARTRIDGE in the DRAWER will display 3 To display data for preceding CARTRIDGES used in any DRAWER a Select the DRAWER b Select the film Lot Number from the Lot drop down list ...

Page 26: ...are Data Film Calibration 2 Select the SUPPLY DRAWER you want to check 3 Click Get Records Plot A DLogE Curve of the characteristics of the film in the selected DRAWER will display The curve is plotted from data stored from the last calibration for this film 4 To make a copy of the DLogE Curve to your LAPTOP click Export Data 5 Copy the file to a folder on the HARD DRIVE of your LAPTOP COMPUTER ...

Page 27: ... Select Diagnostics Diag Summary 2 To view the status of any diagnostic that is now running select Current Status If a diagnostics test is running the name of the test and the Insurance Number of the user will display 3 To display the last results of any diagnostic test select the Test Results tab ...

Page 28: ... From the Diagnostic drop down menu select the test you want 5 To display a Summary of the diagnostic click View 6 If you want to copy the test summary to the HARD DRIVE of your LAPTOP click Save 7 Store the file in a folder on the HARD DRIVE of your LAPTOP COMPUTER ...

Page 29: ... 30JUL07 8F2919 Page 29 of 326 LOCAL PANEL 1 Select Diagnostics MCS Local Panel 2 Click Run The SPEAKER in the LOCAL PANEL should beep This indicates The DRE can communicate with the LOCAL PANEL The SPEAKER in the LOCAL PANEL operates ...

Page 30: ...elect Diagnostics Network 2 To do a Ping test to check that the IMAGER can contact other devices on the network a Enter the IP Address of a device on the network b Use the displayed defaults for Count and Time Out c Click Run d Check for Status and a Summary of the test results Note A Ping test can also be run from the Service Menu of the LOCAL PANEL ...

Page 31: ...age 31 of 326 3 To run a DICOM Echo test to check if the IMAGER can do communication with a another device on the network select the DICOM Echo tab 4 For the device type IP Address Port AE Title 5 Click Run 6 Check for Status and a Summary of the test ...

Page 32: ...DIAGNOSTICS Using the Diagnostics 30JUL07 8F2919 Page 32 of 326 Temperature of the PROCESSOR 1 To check the temperatures of the PROCESSOR zones select Monitoring and Control Temperature Monitor ...

Page 33: ... DRE Test Print 2 Select the parameters for the DRE Test Print a In the Study Label window enter a name up to 20 characters for the print b From the Test Print drop down menu select the type of test print c Use the other drop down menus to configure the test print image d Select the DRAWER you want to print from and the number of copies 3 Click Run ...

Page 34: ...n the MCS MASTER CPU starts 3 tracks of diagnostics In 1 track diagnostics are run for the closed SUPPLY DRAWERS 1 at a time until the test has run for all DRAWERS In the other 2 tracks the MCS starts diagnostics for the remaining MICROCONTROLLERS in the system The tests in the 3 tracks are run at the same time The self tests and calibration run by the system are MCS Diagnostics MCS Clear Films PR...

Page 35: ...GNOSTICS Using the Diagnostics 30JUL07 8F2919 Page 35 of 326 Jam Areas The following are the areas indicated by the film jam errors K K K K K K K K Knob Area 1L Area 1M Area 1U Area 2 Area 3 Area 4 Area 5 ...

Page 36: ...s A PRINTER Status Message and a PRINTER STATUS INFO message are returned to the requesting SERVICE CLASS USER SCU in response to a DICOM PRINTER N GET status request Every error has an associated DICOM PRINTER STATUS INFO message some errors have the same PRINTER STATUS INFO message If no errors occur when a PRINTER N GET request is received a PRINTER STATUS INFO message of NORMAL is returned If ...

Page 37: ...COM PRINTER Status Message DICOM PRINTER Status Information Message FAILURE Condition PRINTER DOWN ELEC DOWN PROC DOWN WARNING Condition BAD SUPPLY MGZ CALIBRATION ERR CHECK PRINTER COVER OPEN EMPTY MEDIASZ MEDIATP FILM JAM FILM TRANS ERR PROC INIT PRINTER INIT PRINTER OFFLINE ...

Page 38: ... mode on Touch screen to wake The next scheduled time might also appear Power Save Mode WARNING PROC INIT Power to the PROCESSOR is OFF Failed Failed See error codes An error has occurred the IMAGER in not able to print films Imager is unlocked Open a door Unlocked WARNING PRINTER OFFLINE The IMAGER is ready to be opened Printing Printing NORMAL NORMAL The IMAGER is able to print films Printing Di...

Page 39: ...E PRINTER DOWN Completing prints in process When shutdown is asked for no more associations are accepted and existing associations are aborted Sleep mode on Touch screen to wake The next scheduled time might also appear Sleep Mode WARNING PRINTER OFFLINE Power to the PROCESSOR is OFF Sorter Open Sorter Open WARNING COVER OPEN The SORTER is open The PRINTER is not ready System Shutting Down Shutdow...

Page 40: ...6 Unlocking please wait IMAGER unlock requested NA The PRINTER is completing prints in process Warming Warming WARNING PROC INIT The PROCESSOR is warming up and will not be ready for the number of minutes displayed LOCAL PANEL PRINTER STATUS DICOM PRINTER STATUS INFO Notes ...

Page 41: ... process for this FILM CARTRIDGE Failed Failed An error has occurred No Cartridge No Cartridge No FILM CARTRIDGE in this FILM SUPPLY DRAWER Open Open The FILM SUPPLY DRAWER is open Please Wait Requires Calibration This or another FILM CARTRIDGE must go through calibration before the IMAGER can print from this FILM SUPPLY DRAWER Calibration does not need user action display normal TRAY information ...

Page 42: ... ID number and an error ID number This number system keeps the error ID numbers given to software components in other IMAGERS for similar errors Note On the LOCAL PANEL SUBSYSTEM ID numbers and error numbers have a hyphen between the numbers but not in the error log For example the LOCAL PANEL displays 20 209 and the error log displays 20209 LOCAL PANEL Message JOB MANAGER Status Notes Imager Offl...

Page 43: ...Y RECLAMATION SUBSYSTEM DELIVERY SUBSYSTEM MIM DATA COMPONENTS SUBSYSTEM MPDB UPDATER SUBSYSTEM VERSION SUBSYSTEM 01 xxx 09 xxx 10 12 13 DRE MCS Interface SERVER LOCAL PANEL SERVICE TOOL 10 xxx 13 xxx 20 21 22 23 26 27 28 29 30 32 33 34 35 MCS General MCS System UPPER FILM SUPPLY DRAWER MIDDLE FILM SUPPLY DRAWER LOWER FILM SUPPLY DRAWER FILM TRANSPORT OPTICS PROCESSOR DENSITOMETER SORTER FILM REGI...

Page 44: ... Software init failure Network will not start up Database failure Yes and an audible alarm Power cycle is necessary Service might be necessary Cycle power and call Service if error returns 1 Alert A failure occurred but it should be temporary and can be repaired by the user Film jam Internal temperature errors Yes and an audible alarm The IMAGER should be able to continue after the user fixes the ...

Page 45: ...n event occurred that should notify the user Preventive maintenance Yes but no audible alarm No recovery necessary 3 Notify An event occurred but it was automatically recovered Communication retries Film pickup retry Approaching warning sidebands or limits of adjustment No A message is written to the log The IMAGER should be able to continue automatically SEVERITY DESCRIPTION EXAMPLE LOCAL PANEL N...

Page 46: ... non specific error has occurred in the CONFIGURATION AND CONTROL SUBSYSTEM Result of Error Unknown This is an unexpected or unknown error and it is not possible to document specific details User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm General Error in Configuration and Control Subsystem NORMAL 3 Notify ...

Page 47: ... and restart it This error means that CONFIGURATION AND CONTROL has initiated more than the number of allowed restarts Result of Error The system will continue to run normally unless the subsystem or component in question cannot run any longer User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Maximum Number of Restarts Exceeded for xxxx xx...

Page 48: ...n An error occurred when reading the CONFIGURATION AND CONTROL SUBSYSTEM configuration information Result of Error The CONFIGURATION AND CONTROL SUBSYSTEM might not execute correctly User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Error Reading Configuration Information NORMAL 2 Warning ...

Page 49: ...en attempting to start a subsystem Result of Error The subsystem in question will not be started User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds Error starting subsystem xxxx xxxx is the subsystem name NORMAL 1 Alert ...

Page 50: ...iption An error occurred when attempting to stop a subsystem Result of Error The subsystem in question might be running Full shutdown will not occur User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Error stopping subsystem xxxx xxxx is the subsystem name NORMAL 1 Alert ...

Page 51: ...t of Error The CONFIGURATION AND CONTROL SUBSYSTEM will attempt to start communications with the subsystem in question If communications is lost the subsystem will be shutdown and restarted User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Configuration and Control Subsystem cannot communicate with subsystem xxxx xxxx is the subsystem name...

Page 52: ...ges Description All subsystems to be started by the CONFIGURATION AND CONTROL SUBSYSTEM have started Result of Error None the system will function normally User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm All subsystems successfully started NORMAL 3 Notify ...

Page 53: ...t Be Shut Down Messages Description The OPERATING SYSTEM did not complete shutting down in the expected time Result of Error None the system will function normally User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm OS could not be shut down NORMAL 3 Notify ...

Page 54: ...UBSYSTEM terminated abnormally Result of Error The subsystem will not be available until it is restarted by the CONFIGURATION AND CONTROL SUBSYSTEM The subsystem might cause associations to be rejected or printing to be delayed User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Subsystem xxxx terminated xxxx is the subsystem name NORMAL 3 N...

Page 55: ...al non specific error has occurred in the DICOM INPUT PRINT SCP SUBSYSTEM Result of Error Unknown This is an unexpected or unknown error and it is not possible to document specific details User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm General Error in DICOM Input Print SCP Subsystem NORMAL 3 Notify ...

Page 56: ...t of Error The SCU will not be able to open an association and will not be able to deliver a print This is possibly a transient problem The user cannot remedy this from the IMAGER The DICOM result reason codes available are 2 rejected transient 3 local limit exceeded User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Maximum Number of Assoc...

Page 57: ... SCU will not be able to open an association and will not be able to deliver a print job This is a DICOM SCU configuration problem not an operational error The user cannot remedy this from the IMAGER The DICOM result reason codes available are 1 rejected permanent 7 AE title not recognized User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm ...

Page 58: ...the presentation context items were successfully negotiated This might occur for example when a DICOM STORE SCU tries to communicate with a DICOM PRINT SCP Result of Error The SCU will not be able to open an association and will not be able to deliver a print job This is a DICOM SCU configuration problem not an operational error The user cannot remedy this from the IMAGER The DICOM result reason c...

Page 59: ...is full Result of Error If a study is in progress it cannot be completed New studies will not be started User Action 1 Delete jobs in the unprintable jobs queue 2 Load the correct film type for jobs in the waiting for media queue 3 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Disk Full Alarm sounds Disk Full NORMAL 1 Alert ...

Page 60: ... maintain a TCP IP connection with an SCU to which the SCU should reply This error means that the SCU did not respond to the SCP s Are You There packet Result of Error The SCP will abort the association tied to this connection User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Send Keep Alive Message Failed NORMAL 2 Warning ...

Page 61: ... Lost Messages Description The SCP has lost the network connection to an SCU Result of Error The SCP will abort the association tied to this connection User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Network lost connection NORMAL 2 Warning ...

Page 62: ...n the FILM BOX or IMAGE BOX is not present Result of Error The MIM CORE will not process the print job This is a DICOM SCU configuration problem not an operational error The user cannot remedy this from the IMAGER User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Referencing PLUT Missing NORMAL 1 Alert ...

Page 63: ... If the SOP is a FILM SESSION or FILM BOX the pages will not be printed This is a DICOM SCU configuration problem not an operational error The user cannot remedy this from the IMAGER User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Required attribute wwww xxxx yyyy is missing from zzzz SOP wwww xxxx yyyy is the tag name and number zzzz is...

Page 64: ...he SOP is a FILM SESSION or FILM BOX the pages will not be printed This is a DICOM SCU configuration problem not an operational error The user cannot remedy this from the IMAGER User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Invalid attribute wwww xxxx yyyy encountered in zzzz SOP wwww xxxx yyyy is the tag name and number zzzz is the SO...

Page 65: ...n An unknown error has occurred in the STUDY RECLAMATION SUBSYSTEM Result of Error Unknown This is an unexpected or unknown error and it is not possible to document specific details User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Error Unknown NORMAL 2 Warning ...

Page 66: ...Description An error occurred reading the STUDY RECLAMATION SUBSYSTEM configuration information Result of Error The STUDY RECLAMATION SUBSYSTEM might not execute correctly User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Error reading configuration information NORMAL 2 Warning ...

Page 67: ...r the high water mark is reached or the scheduled time for study reclamation has occurred Result of Error Completed studies will be removed from the system until the low water mark is reached User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm High water mark reached Scheduled study reclamation time reached NORMAL 3 Notify ...

Page 68: ...RY SUBSYSTEM could not reach the low water mark when running the STORAGE RECLAMATION ROUTINES Result of Error None The MIM CORE should continue to run normally This warning might continue to be logged User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Low water mark could not be reached NORMAL 3 Notify ...

Page 69: ...ater Mark Reached Messages Description Study reclamation has is completed because the low water mark was successfully reached Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Low water mark reached NORMAL 3 Notify ...

Page 70: ...ON SUBSYSTEM has received a shutdown request Result of Error None Possibly the system is shutting down Also it might be that a configuration information change that needs the subsystem to be restarted User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Study reclamation is shutting down NORMAL 3 Notify ...

Page 71: ...A general non specific error has occurred in the DELIVERY SUBSYSTEM Result of Error Unknown This is an unexpected or unknown error and it is not possible to document specific details User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm General error in delivery subsystem NORMAL 3 Notify ...

Page 72: ... the DELIVERY SUBSYSTEM Result of Error Jobs will not be completed as the software interface for the DELIVERY SUBSYSTEM could not be started User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds General start up error NORMAL 1 Alert ...

Page 73: ...ding the DELIVERY SUBSYSTEM configuration information Result of Error The DELIVERY SUBSYSTEM might not execute correctly User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds Bad or missing configuration information NORMAL 1 Alert ...

Page 74: ...n internal error User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds General job processing error 1 Start up error 2 Bad or missing configuration data 3 Resource allocation error 4 Mode switching error 5 Multi threading error 6 Message exchange error 7 Status call...

Page 75: ...dy Messages Description The PRINTER is not ready to receive delivery jobs Result of Error Jobs will not be completed when the PRINTER is in this state User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Printer not in ready state NORMAL 2 Warning ...

Page 76: ...figuration information Result of Error The job will not be completed User Action 1 Delete jobs in the unprintable jobs queue 2 Restart the LASER IMAGER from the LOCAL PANEL 3 Send another print job from the image source 4 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds Invalid media NORMAL 1 Alert ...

Page 77: ...26 Error 06 104 Media Not Available Messages Description The media is not available Result of Error The job will not be completed User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Media unavailable NORMAL 2 Warning ...

Page 78: ...06 105 Page Delivery Success Messages Description A page was successfully completed This is not an error condition Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Page delivery success NORMAL 3 Notify ...

Page 79: ... 06 106 Job Delivery Success Messages Description A job was successfully completed This is not an error condition Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Job delivery success NORMAL 3 Notify ...

Page 80: ...ing to access the MIM CORE DATABASE The DATABASE might be corrupted or a timeout occurred Result of Error The job will not be completed User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds Database access error NORMAL 1 Alert ...

Page 81: ...he DESTINATION BUSINESS LOGIC component The DATABASE might be corrupted or a timeout occurred Result of Error The job will not be completed User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds Database access error NORMAL 1 Alert ...

Page 82: ... images at lower level Error initializing text line lists or variables Result of Error The page will not be completed User Action 1 Delete jobs in the unprintable jobs queue 2 Restart the LASER IMAGER from the LOCAL PANEL 3 Send another print job from the image source 4 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Image Page Error Alarm sounds General pa...

Page 83: ... because of an internal error User Action 1 Delete jobs in the unprintable jobs queue 2 Restart the LASER IMAGER from the LOCAL PANEL LOCAL PANEL LOG DICOM SEVERITY MIM Core Image Rendering Error Alarm sounds 1 General error rendering the page 2 Error after completion of chain forming 3 Error rendering trim or rectangle 4 Error rendering text xxxx xxxx is the text to be rendered 5 Error rendering ...

Page 84: ...DIAGNOSTICS Status Messages and Error Codes 30JUL07 8F2919 Page 84 of 326 3 Send another print job from the image source 4 If the error occurs again call Service FE Action ...

Page 85: ...ll not be processed because it could not be rendered because of an internal error User Action 1 Delete jobs in the unprintable jobs queue 2 Restart the LASER IMAGER from the LOCAL PANEL 3 Send another print job from the image source 4 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Image Data Error Alarm sounds 1 Width and or height less than zero 2 Bits be...

Page 86: ... Requested Messages Description The PAGE BUILDER component is asked to stop a rendering that is in progress Result of Error The page will not be processed because rendering was stopped User Action None FE Action LOCAL PANEL LOG DICOM SEVERITY None No alarm Render abort requested NORMAL 1 Alert ...

Page 87: ...oftware Error Alarm sounds 1 Page width and or height are less than zero 2 Invalid page band 3 Invalid coordinates for indicator icon x1 y1 x2 y2 x and y are the calculated icon coordinates 4 Image size mismatch error 5 Error in cropping coordinates 6 The file xxxx could not be found xxxx is the name of the image file 7 Error reading image xxxx xxxx is the name of the image file 8 General error ac...

Page 88: ...he page will not be processed because it could not be rendered because of an internal error User Action 1 Delete jobs in the unprintable jobs queue 2 Restart the LASER IMAGER from the LOCAL PANEL 3 Send another print job from the image source 4 If the error occurs again call Service FE Action ...

Page 89: ...n the unprintable jobs queue 2 Restart the LASER IMAGER from the LOCAL PANEL 3 Send another print job from the image source 4 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds 1 General error building image processing chain 2 Error in tonescaling xxxx xxxx is the specific tonescaling problem 3 Error forming the color proce...

Page 90: ...cause of an internal error communicating with the MIS This is an error on an individual job and does not affect PRINTER operation The job will go to the undelivered queue User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm 1 BIND error Code xx Page ID yyy 2 STAGE error Code xx Page ID yyy 3 PRINT error Code xx Page ID yyy 4 EXPOSE error Code...

Page 91: ...es Description The requested media is not loaded in the PRINTER Result of Error The job will not be completed until the media is loaded User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Media xxxx not loaded xxxx is the requested media NORMAL 2 Warning ...

Page 92: ... Messages Description The media is not calibrated Result of Error The job will not be completed until the media is calibrated User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Media xxxx requires calibration xxxx is the requested media NORMAL 2 Warning ...

Page 93: ... Messages Description The media is calibrating Result of Error Jobs will not be completed until the calibration cycle is complete User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Media xxxx is calibrating xxxx is the requested media NORMAL 3 Notify ...

Page 94: ...n An error occurred during media calibration Result of Error Jobs might not be completed because the media could not be calibrated User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Error calibrating media xxxx xxxx is the media that failed calibration NORMAL 3 Notify ...

Page 95: ...is Empty Messages Description The media is empty Result of Error The job will not be completed until the media is loaded User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Media xxxx is empty xxxx is the requested media NORMAL 2 Warning ...

Page 96: ...6 506 Jam in Media Messages Description There is a jam in the media Result of Error The job will not be completed until the media jam is cleared User Action None FE Action LOCAL PANEL LOG DICOM SEVERITY None No alarm Media xxxx is jammed xxxx is the requested media NORMAL 2 Warning ...

Page 97: ... in the DELIVERY SUBSYSTEM Result of Error Unknown This is an unexpected or unknown error and it is not possible to document specific details User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIM Core Internal Software Error Alarm sounds General error in delivery subsystem NORMAL 1 Alert ...

Page 98: ...l memory failures object creations failing system call failures Result of Error The DRE is not functional User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 Make a copy of the log with the error 2 Send the log to Service Engineering 3 Make a copy of the system configuration 4 Load new system software 5 Restore the system configuration L...

Page 99: ...BSYSTEM has failed Result of Error The DRE is not functional User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 Make a copy of the log with the error 2 Send the log to Service Engineering 3 Make a copy of the system configuration 4 Load new system software 5 Restore the system configuration LOCAL PANEL LOG DICOM SEVERITY MIS Image Buffe...

Page 100: ...iled Result of Error The PRINTER can receive jobs and print but cannot track total print counts prints to PM prints to FILTER change or the history of CARTRIDGE and calibration data User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 Make a copy of the log with the error 2 Send the log to Service Engineering 3 Make a copy of the system c...

Page 101: ...hat will result in a service call Result of Error The PRINTER cannot print films The SPOOLER is able to receive jobs but new films will not be initiated Films in progress will not be completed User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action LOCAL PANEL LOG DICOM SEVERITY MIS MCS Communications Failure Alarm sounds MIS to MCS Communicat...

Page 102: ...OCAL PANEL SUBSYSTEM The log message will have additional details Result of Error The LOCAL PANEL might not function correctly but it can continue operating It is possible that a specific functionality is not available User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Local Panel Internal Error NORMAL 1 Alert ...

Page 103: ...ed The error is logged and it cannot be cleared until the system is re ghosted Result of Error The system is unusable because the failed software update could have left the system in any state It is unknown if diagnostic quality images can be generated Images will not be printed User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Software Up...

Page 104: ...ountered when the LOCAL PANEL is attempting to use it The log message will have additional details Result of Error The LOCAL PANEL cannot interact with the system FLASH The LOCAL PANEL can continue operating It is possible that backup restore is not available User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Local Panel Flash Access Error ...

Page 105: ...L SUBSYSTEM has failed This could be continual memory failures object creations failing system call failures COMMON SERVICE PLATFORM CSP errors Result of Error Service operations might not be operating correctly The DRE should be functional but might not respond correctly to the service function or the function might fail User Action None FE Action None this is a log notification only LOCAL PANEL ...

Page 106: ...OOL SUBSYSTEM has failed This could be database read write or update failed Result of Error Service operations might not be operating correctly particularly diagnostic operations The DRE should be functional but might not respond correctly to the service function or the function might fail User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm ...

Page 107: ...d on the IMAGER Result of Error The DRE is not functional User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER 2 Check the connection on the DATAPATH USB CABLE 3 Install a new HARD DRIVE See...

Page 108: ...m will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER 2 Re ghost the DRE SOFTWARE 3 Install a new HARD DRIVE See the ADJUSTMENTS AND REPL...

Page 109: ... films will not be initiated Films in progress will not be completed User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER 2 Use the SERVICE TOOL to check that the version of software in each...

Page 110: ...Error Codes 30JUL07 8F2919 Page 110 of 326 5 Install a new MICRO BOARD 6 Install a new HARD DRIVE See the ADJUSTMENTS AND REPLACEMENTS HARD DRIVE DRE AY 7 Install a new DRE See the ADJUSTMENTS AND REPLACEMENTS DICOM RASTER ENGINE DRE ...

Page 111: ... The SPOOLER is able to receive jobs but new films will not be initiated Films in progress will not be completed User Action 1 Close all open DRAWERS and DOORS 2 Restart the LASER IMAGER from the LOCAL PANEL 3 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 Close the open DOOR or DRAWER 2 ...

Page 112: ...CH POSITION SENSOR See the ILLUSTRATED PARTS LIST MAIN AY Side Components LATCH SPRING Install a new SPRING See the ILLUSTRATED PARTS LIST MAIN AY Top Components LATCH MOTOR Install a new LATCH MOTOR BRACKET AY See the ILLUSTRATED PARTS LIST MAIN AY Side Components BLINDMATE CONNECTOR Adjust the BLINDMATE See the ADJUSTMENTS AND REPLACEMENTS BLINDMATE BRACKET ...

Page 113: ...ssage is written to the log when the sheet count exceeds the PM sheet count limit An indicator on the LOCAL PANEL is used to notify the operator of the PM requirement Result of Error None User Action 1 Notify Service that PM is necessary FE Action 1 Do the PM LOCAL PANEL LOG DICOM SEVERITY None No Alarm Preventive Maintenance Required NORMAL 3 Notify ...

Page 114: ...hen the sheet count exceeds the sheet count limit for the Preventive Maintenance Module PMM An indicator on the LOCAL PANEL is used to notify the operator of the PM requirement Result of Error None User Action 1 If at 15 000 45 000 75 000 etc Replace PMM MODULE FE Action 1 Change the PMM LOCAL PANEL LOG DICOM SEVERITY None No Alarm Processor Filter Maintenance Required NORMAL 3 Notify ...

Page 115: ...e The faulty image will be printed The job will be failed placed in the undelivered queue and rescheduled for delivery Note Error 20 915 is generated when this problem is detected on 3 successive films Result of Error None the system will function normally User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No Alarm Image transfer failure Wrong numbe...

Page 116: ... to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films must be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the L...

Page 117: ...s and Error Codes 30JUL07 8F2919 Page 117 of 326 4 Install a new DRE See the ADJUSTMENTS AND REPLACEMENTS DICOM RASTER ENGINE DRE 5 Install a new OPTICS AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY ...

Page 118: ...OSURE TRANSPORT SUBSYSTEM A faulty image is printed on each failure The job will be failed placed in the undelivered queue and rescheduled for delivery Note Error 20 918 is generated when this problem is detected on 3 successive films Result of Error None the system will function normally User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No Alarm I...

Page 119: ... print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1...

Page 120: ...s and Error Codes 30JUL07 8F2919 Page 120 of 326 4 Install a new DRE See the ADJUSTMENTS AND REPLACEMENTS DICOM RASTER ENGINE DRE 5 Install a new OPTICS AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY ...

Page 121: ...will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER 2 Install a new DATAPATH USB CABLE from the DRE to the OPTICS MODULE 3 Install a new ...

Page 122: ...DIAGNOSTICS Status Messages and Error Codes 30JUL07 8F2919 Page 122 of 326 5 Install a new OPTICS AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY ...

Page 123: ...up sequence must complete successfully for this error to occur The error is cleared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot make a print from the SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Open the DRAWER 2 Close the CARTRIDGE to prevent film fog 3 ...

Page 124: ...ICKUP AY ROLLERS and DRIVE Section BLINDMATE CONNECTOR Adjust THE BLINDMATE See the ADJUSTMENTS AND REPLACEMENTS BLINDMATE BRACKET PICKUP FILM AT FEED SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR FILM AT FEED S5 PICKUP AY PICKUP FEED ROLLER OPEN CLOSE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR FEED ROLLE...

Page 125: ...w SPRING See the ILLUSTRATED PARTS LIST PICKUP AY ROLLERS and DRIVE Section WIRING HARNESS Install a new HARNESS See the ILLUSTRATED PARTS LIST PICKUP AY HARNESS and TUBING FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FSB AY PICKUP AY SUCTION CUPS Install new SUCTION CUPS See the ADJUSTMENTS AND REPLACEMENTS SUCTION CUPS PICKUP AY CUP CARRIAGE Check that the CUP CAR...

Page 126: ... place it into feed position on the first or second attempt The error is detected by the FILM AT FEED SENSOR S5x when the leading edge of the film does not reach the SENSOR Note Error 2x 116 is generated when this problem is detected on 3 successive failures Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No Alarm xxxxx Suppl...

Page 127: ...leared when the IMAGER power is cycled Result of Error The system cannot make a print from the SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue w...

Page 128: ...NSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CUPS ENGAGED S2 and PICKUP HOME S1 PICKUP AY ROLLER DRIVE MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR FEED ROLLER DRIVE M2 PICKUP AY ROLLER DRIVE MOTOR BELT Install a new BELTS See the ADJUSTMENTS AND REPLACEMENTS BELTS ROLLER DRIVE and ROLLER OPEN CLOSE PICKUP AY PICKUP DRIVE MOTOR ...

Page 129: ...e PICKUP DRIVE MOTOR M3 was operating long enough to leave home the most possible cause is a failed CUPS ENGAGED SENSOR S2x A FILM OUT SENSOR S3x that failed in the unblocked state might also be a cause The error is cleared when the IMAGER power is cycled Result of Error The system cannot make a print from the SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other ava...

Page 130: ...NTS BLINDMATE BRACKET FILM OUT SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR FILM OUT S3 PICKUP AY CUPS ENGAGED SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CUPS ENGAGED S2 and PICKUP HOME S1 PICKUP AY ROLLER DRIVE MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR FEED ROLLER DRIVE M2 ...

Page 131: ...l a new PICKUP AY See the ADJUSTMENTS AND REPLACEMENTS PICKUP AY WIRING HARNESS Install a new HARNESS See the ILLUSTRATED PARTS LIST PICKUP AY HARNESS and TUBING FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FSB AY PICKUP AY Check If an existing problem ...

Page 132: ...d the CUPS ENGAGED SENSOR S2x so it is possible that one of these components failed or the film pack is blocked stuck together preventing the movement of the PICKUP VACUUM ARM The error is cleared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot make a print from the SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print ...

Page 133: ...D REPLACEMENTS BLINDMATE BRACKET CUPS ENGAGED SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CUPS ENGAGED S2 and PICKUP HOME S1 PICKUP AY ROLLER DRIVE MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR FEED ROLLER DRIVE M2 PICKUP AY ROLLER DRIVE MOTOR BELT Install a new BELTS See the ADJUSTMENTS AND REPLACEMENTS BELTS ROLLER DRIVE and...

Page 134: ...ot detect the motion The error is cleared when the IMAGER is power cycled Result of Error The system cannot make a print from the SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action prov...

Page 135: ...IDGE CONNECTORS and CABLES Seat all CONNECTORS FEED ROLLER POSITION SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR FEED ROLLER POSITION S4 PICKUP AY ROLLER DRIVE MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR FEED ROLLER DRIVE M2 PICKUP AY PICKUP DRIVE MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR DRIVE M3 P...

Page 136: ...e center or the trailing edge of the film The error is cleared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be started Films downstream of the subsystem will be completed All other films must be reprinted User Action 1 Open the DRAWER 2 Close the CARTR...

Page 137: ...Remove films For an obstruction Remove the obstruction CONNECTORS and CABLES Seat all CONNECTORS ROLLERS Install new ROLLERS FILM AT FEED SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR FILM AT FEED S5 PICKUP AY FEED ROLLER OPEN CLOSE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR FEED ROLLER OPEN CLOSE M1 PICK...

Page 138: ...leared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot make a print from the SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a ...

Page 139: ...DMATE CONNECTOR Adjust THE BLINDMATE See the ADJUSTMENTS AND REPLACEMENTS BLINDMATE BRACKET RF TAG ANTENNA CABLE 1 Seat the CONNECTOR 2 Install a new HARNESS See the ADJUSTMENTS AND REPLACEMENTS HARNESS ROLLBACK AY RF ANTENNA BOARD Install a new BOARD See the ADJUSTMENTS AND REPLACEMENTS RF ANTENNA BOARD AY ROLLBACK AY FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FS...

Page 140: ... CR mark information The error is cleared when the SUPPLY DRAWER is opened and closed as indicated by SUPPLY DRAWER INTERLOCK Result of Error The system cannot print from the UPPER SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Service FE Action Do the actions b...

Page 141: ...DMATE CONNECTOR Adjust THE BLINDMATE See the ADJUSTMENTS AND REPLACEMENTS BLINDMATE BRACKET RF TAG ANTENNA CABLE 1 Seat the CONNECTOR 2 Install a new HARNESS See the ADJUSTMENTS AND REPLACEMENTS HARNESS ROLLBACK AY RF ANTENNA BOARD Install a new BOARD See the ADJUSTMENTS AND REPLACEMENTS RF ANTENNA BOARD AY ROLLBACK AY FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FS...

Page 142: ...Y DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue ...

Page 143: ...DMATE CONNECTOR Adjust THE BLINDMATE See the ADJUSTMENTS AND REPLACEMENTS BLINDMATE BRACKET RF TAG ANTENNA CABLE 1 Seat the CONNECTOR 2 Install a new HARNESS See the ADJUSTMENTS AND REPLACEMENTS HARNESS ROLLBACK AY RF ANTENNA BOARD Install a new BOARD See the ADJUSTMENTS AND REPLACEMENTS RF ANTENNA BOARD AY ROLLBACK AY FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FS...

Page 144: ...of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize th...

Page 145: ...DMATE CONNECTOR Adjust THE BLINDMATE See the ADJUSTMENTS AND REPLACEMENTS BLINDMATE BRACKET RF TAG ANTENNA CABLE 1 Seat the CONNECTOR 2 Install a new HARNESS See the ADJUSTMENTS AND REPLACEMENTS HARNESS ROLLBACK AY RF ANTENNA BOARD Install a new BOARD See the ADJUSTMENTS AND REPLACEMENTS RF ANTENNA BOARD AY ROLLBACK AY FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FS...

Page 146: ...correctly installed The error is cleared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Service FE Action Do the actions below in seque...

Page 147: ...SOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CARTRIDGE PRESENCE S8 ROLLBACK AY SENSOR FLAG Install a new FLAG See the ADJUSTMENTS AND REPLACEMENTS FLAG ROLLBACK AY MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR AY M6 ROLLBACK AY ROLLER NEST for obstructions or damage Install a new NEST See the ILLUSTRATED PARTS LIST ROLLBACK AY Overview ROLLER GEAR Ins...

Page 148: ...s possible that one of the 2 SENSORS failed The error is cleared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Service FE Action Do th...

Page 149: ...ACK AY ROLLBACK OPEN SENSOR use the SERVICE TOOL Install a new SENSOR OPEN S7 ROLLBACK AY SENSOR FLAG Install a new FLAG See the ADJUSTMENTS AND REPLACEMENTS FLAG ROLLBACK AY MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR AY M6 ROLLBACK AY ROLLER GEAR Install a new GEAR See the ILLUSTRATED PARTS LIST ROLLBACK AY Overview LATCH mechanism Install a new LATC...

Page 150: ...CK HOME SENSOR S6x but this error can also be caused by failure of the ROLLBACK OPEN SENSOR S7x or by failure of the ROLLBACK MOTOR M6x The error is cleared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK or when power is cycled on the IMAGER Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any...

Page 151: ...MATE BRACKET ROLLBACK HOME SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR HOME S6 ROLLBACK AY ROLLBACK OPEN SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR OPEN S7 ROLLBACK AY SENSOR FLAG Install a new FLAG See the ADJUSTMENTS AND REPLACEMENTS FLAG ROLLBACK AY MOTOR use the SERVICE TOOL Install a new MOTOR S...

Page 152: ... See the ADJUSTMENTS AND REPLACEMENTS ROLLBACK AY CHAIN for obstructions or damage 1 Remove obstruction 2 Install a new CHAIN See the ADJUSTMENTS AND REPLACEMENTS CHAIN ROLLBACK AY FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FSB AY PICKUP AY Check If an existing problem ...

Page 153: ...or both SENSORS failed The error is cleared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK or when power is cycled on the IMAGER Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Servic...

Page 154: ...w SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR OPEN S7 ROLLBACK AY SENSOR FLAG Install a new FLAG See the ADJUSTMENTS AND REPLACEMENTS FLAG ROLLBACK AY MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR AY M6 ROLLBACK AY ROLLER GEAR Install a new GEAR See the ILLUSTRATED PARTS LIST ROLLBACK AY Overview LATCH mechanism Install a new LATCH See the ILLUSTR...

Page 155: ...DIAGNOSTICS Status Messages and Error Codes 30JUL07 8F2919 Page 155 of 326 3 Install a new ROLLBACK AY See the ADJUSTMENTS AND REPLACEMENTS ROLLBACK AY ...

Page 156: ...ROLLBACK closed This is possibly a failure of the ROLLBACK OPEN SENSOR S7x but this error can also be caused by a ROLLBACK MOTOR M6x failure The error is cleared when the SUPPLY DRAWER is closed as indicated by the SUPPLY DRAWER INTERLOCK or when power is cycled on the IMAGER Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print fro...

Page 157: ...MATE BRACKET ROLLBACK HOME SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR HOME S6 ROLLBACK AY ROLLBACK OPEN SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR OPEN S7 ROLLBACK AY SENSOR FLAG Install a new FLAG See the ADJUSTMENTS AND REPLACEMENTS FLAG ROLLBACK AY MOTOR use the SERVICE TOOL Install a new MOTOR S...

Page 158: ... See the ADJUSTMENTS AND REPLACEMENTS ROLLBACK AY CHAIN for obstructions or damage 1 Remove obstruction 2 Install a new CHAIN See the ADJUSTMENTS AND REPLACEMENTS CHAIN ROLLBACK AY FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FSB AY PICKUP AY Check If an existing problem ...

Page 159: ...DOORS are opened then closed Open is indicated by the I3 INTERLOCK PROCESSOR and vertical DOOR open Close is indicated by the INTERLOCK closed Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Open all DOORS and the FILM ...

Page 160: ...isting problem For film jam in area 1 Remove films FILM CARTRIDGE Install a new FILM CARTRIDGE ACCUMULATOR FILM DETECT SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR ACCUMULATOR FILM DETECT S9 FILM REGISTRATION AY REGISTRATION CLAMP SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CLAMP S11 FILM REGISTRATION ...

Page 161: ...r calibration is asked for or when the SUPPLY DRAWER closes as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Service FE Action Do the actions below in sequence If an acti...

Page 162: ...DJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem FILM CARTRIDGE Install a new FILM CARTRIDGE CONNECTORS and CABLES Seat all CONNECTORS BELTS on TURNAROUND Install a new BELT ROLLERS on TURNAROUND Install a new TURNAROUND AY See the ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER ...

Page 163: ...closes as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Service Note The operator can ignore calibration for this FILM CARTRIDGE FE Action Do the actions below in sequenc...

Page 164: ...ng 3 Install a new TURNAROUND AY See the ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem FILM CARTRIDGE Install a new FILM CARTRIDGE CONNECTORS and CABLES Seat all CONNECTORS ROLLERS on TURNAROUND Install a new TURNAROUND AY See the ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER ...

Page 165: ...closes as indicated by the SUPPLY DRAWER INTERLOCK Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Install a new FILM CARTRIDGE 2 If the error occurs again call Service Note The operator can ignore calibration for this FILM CARTRIDGE FE Action Do the actions below in sequenc...

Page 166: ...ng 3 Install a new TURNAROUND AY See the ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem FILM CARTRIDGE Install a new FILM CARTRIDGE CONNECTORS and CABLES Seat all CONNECTORS ROLLERS on TURNAROUND Install a new TURNAROUND AY See the ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER ...

Page 167: ...ed when power is cycled on the IMAGER Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the...

Page 168: ...ck If an existing problem FILM CARTRIDGE Install a new FILM CARTRIDGE RF TAG ANTENNA CABLE 1 Seat the CONNECTOR 2 Install a new HARNESS See the ADJUSTMENTS AND REPLACEMENTS HARNESS ROLLBACK AY RF ANTENNA BOARD Install a new BOARD See the ADJUSTMENTS AND REPLACEMENTS RF ANTENNA BOARD AY ROLLBACK AY ...

Page 169: ...Result of Error The system cannot print from this SUPPLY DRAWER The SPOOLER is able to receive and queue jobs and print from any other available SUPPLY DRAWER User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then ener...

Page 170: ...following Check If an existing problem CONNECTORS and CABLES to from the PDB Seat all CONNECTORS BLINDMATE CONNECTOR Adjust THE BLINDMATE See the ADJUSTMENTS AND REPLACEMENTS BLINDMATE BRACKET FSB Install a new FSB See the ADJUSTMENTS AND REPLACEMENTS FILM SUPPLY BOARD FSB AY PICKUP AY ...

Page 171: ...ccurs when the MCS tries to communicate with the MICROCONTROLLER but receives no response The error is cleared when power is cycled on the IMAGER Note Error 2x 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to commun...

Page 172: ... Close is indicated by the I3 INTERLOCK closed Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Open all DOORS on the IMAGER 2 Remove any film jams in area 2 Note Use the FILM FEED KNOB in area 2 to move the film down to...

Page 173: ... FILM CARTRIDGE Install a new FILM CARTRIDGE BELTS Install new BELTS ACCUMULATOR FILM DETECT SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR ACCUMULATOR FILM DETECT S9 FILM REGISTRATION AY REGISTRATION CLAMP SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CLAMP S11 FILM REGISTRATION AY ROLLER DRIVE MOTOR use ...

Page 174: ...ted by the I3 INTERLOCK closed Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Open all DOORS on the IMAGER 2 Remove any film jams in area 2 Note Use the FILM FEED KNOB in area 2 to move the film down to the bottom acce...

Page 175: ...JUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For film jam in area 2 Remove films FILM CARTRIDGE Install a new FILM CARTRIDGE BELTS Install new BELTS EXPOSURE ENTRANCE SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR ENTRANCE S12 EXPOSURE TRANSPORT AY EXPOSURE ENTRANCE CLAMP MOTOR use the SERVICE TOOL Install...

Page 176: ... PRINTER will be marked bad and reprinted automatically This includes films in the lower part of the PRINTER that will also be cleared automatically when the IMAGER is locked Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Actio...

Page 177: ...en energize the LASER IMAGER 2 Do the following 3 Install a new EXPOSURE TRANSPORT AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For film jam in areas 2 and 3 Remove films FILM CARTRIDGE Install a new FILM CARTRIDGE BELTS Install new BELTS EXPOSURE ENTRANCE SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMEN...

Page 178: ...films in the PRINTER will be marked bad and reprinted automatically This includes films in the lower part of the PRINTER that will also be cleared automatically when the IMAGER is locked Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinte...

Page 179: ...uence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER 2 Do the following 3 Install a new TURNAROUND AY See the ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem For film jam in areas 3 and 4 Remove films FILM CARTRIDGE Install a new FILM CARTRIDGE BELTS Install new BELTS DENSITOMETER FIL...

Page 180: ... the PRINTER will be marked bad and reprinted automatically This includes films in the lower part of the PRINTER that will also be cleared automatically when the IMAGER is locked Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User A...

Page 181: ... the LASER IMAGER 2 Do the following 3 Install a new TURNAROUND AY See the ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem For film jam in area 5 Remove films FILM CARTRIDGE Install a new FILM CARTRIDGE BELTS Install new BELTS DENSITOMETER FILM DETECT SENSOR use the SERVICE TOOL Install a new TURNAROUND AY See the ADJUSTMENTS AND REPLACEMENTS TURNAROUND A...

Page 182: ...n the IMAGER Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a reso...

Page 183: ...ORT AY Check If an existing problem CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS OPTICS ATTENUATOR HOME SENSOR use the SERVICE TOOL Install a new OPTICS AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY ATTENUATOR MOTOR use the SERVICE TOOL Install a new OPTICS AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY ...

Page 184: ...able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize t...

Page 185: ...Y See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS POWER MODULE Install a new POWER MODULE See the ADJUSTMENTS AND REPLACEMENTS POWER MODULE AY ...

Page 186: ...ceive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER I...

Page 187: ...T AY Check If an existing problem CONNECTORS on the PDB ETB SOP Seat the CONNECTORS PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY SOP Install a new SOP See the ADJUSTMENTS AND REPLACEMENTS START OF PAGE SOP BOARD S13 EXPOSURE TRANSPORT AY ...

Page 188: ...POOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then ...

Page 189: ...ENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS SERVICE SWITCH 1 Close the SERVICE SWITCH 2 Install a new SERVICE SWITCH PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 190: ...int films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De...

Page 191: ... AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 192: ... The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize...

Page 193: ... AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 194: ...not print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other action...

Page 195: ... AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 196: ...able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize t...

Page 197: ... AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 198: ... SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize the...

Page 199: ... AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 200: ... SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize the...

Page 201: ... AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 202: ...ption This occurs when the laser dynamic range is adjusted and the value is within 10 of the adjustment range Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Optics Calibration detected laser dynamic range approaching predefined limits NORMAL 3 Notify ...

Page 203: ... Description This occurs if the maximum LASER power measured by OPTICS calibration and diagnostics is approaching the limit Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Optics Calibration detected Optics Maximum Power approaching limits NORMAL 3 Notify ...

Page 204: ...rint films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 D...

Page 205: ... AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem For open DOORS or DRAWERS Close all DOORS and DRAWERS CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 206: ... is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energi...

Page 207: ...DRIVE BELT Install a new BELT See the ILLUSTRATED PARTS LIST EXPOSURE TRANSPORT AY EXPOSURE ENTRANCE CLAMP MOTOR use the SERVICE TOOL Install a new MOTOR ENTRANCE EXIT CLAMP M10 M11 EXPOSURE TRANSPORT AY EXPOSURE TRANSPORT DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR DRIVE M12 EXPOSURE TRANSPORT AY PINCH ROLLERS Install a new ROLLER See the ADJUST...

Page 208: ... SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize the...

Page 209: ...nstall a new BELT See the ILLUSTRATED PARTS LIST EXPOSURE TRANSPORT AY EXPOSURE ENTRANCE CLAMP MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR ENTRANCE EXIT CLAMP M10 M11 EXPOSURE TRANSPORT AY DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR DRIVE M12 EXPOSURE TRANSPORT AY PINCH ROLLERS Install a new ROLLER Se...

Page 210: ...R cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other a...

Page 211: ...all a new OPTICS AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY ...

Page 212: ... receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASE...

Page 213: ...all a new OPTICS AY See the ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY ...

Page 214: ...necessary time The faulty image will be printed The job will be failed placed in the undelivered queue and rescheduled for delivery Note Error 27 650 is generated when this problem is detected on 3 successive films Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Image Exposure timed out job will be reprinted NORMAL 3...

Page 215: ... receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASE...

Page 216: ...REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem CONNECTORS on the DRE USB CABLE PDB ETB SOP BOARD Seat the CONNECTORS ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 217: ...MCS tries to communicate with the MICROCONTROLLER but receives no response Note Error 27 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to communicate with Optics BPM NORMAL 3 Notify ...

Page 218: ...ceive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER I...

Page 219: ...ND REPLACEMENTS THERMAL PROCESSOR AY Extending the PROCESSOR Check If an existing problem CONNECTORS on the DRE USB CABLE PDB PCB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 220: ...t of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not cont...

Page 221: ...ost 4 Install a new PROCESSOR AY See the ADJUSTMENTS AND REPLACEMENTS THERMAL PROCESSOR AY Extending the PROCESSOR Check If an existing problem CONNECTORS on the PDB PCB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 222: ...AGER Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution d...

Page 223: ...UIT BREAKER RTD settings use the SERVICE TOOL Install a new PROCESSOR AY See the ADJUSTMENTS AND REPLACEMENTS THERMAL PROCESSOR AY Extending the PROCESSOR DRUM Install a new DRUM See the ADJUSTMENTS AND REPLACEMENTS DRUM AY THERMAL PROCESSOR AY PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND R...

Page 224: ...esult of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not ...

Page 225: ...UIT BREAKER RTD settings use the SERVICE TOOL Install a new PROCESSOR AY See the ADJUSTMENTS AND REPLACEMENTS THERMAL PROCESSOR AY Extending the PROCESSOR DRUM Install a new DRUM See the ADJUSTMENTS AND REPLACEMENTS DRUM AY THERMAL PROCESSOR AY PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND R...

Page 226: ...films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De ene...

Page 227: ...UIT BREAKER RTD settings use the SERVICE TOOL Install a new PROCESSOR AY See the ADJUSTMENTS AND REPLACEMENTS THERMAL PROCESSOR AY Extending the PROCESSOR DRUM Install a new DRUM See the ADJUSTMENTS AND REPLACEMENTS DRUM AY THERMAL PROCESSOR AY PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND R...

Page 228: ...sult of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not c...

Page 229: ...CTORS CIRCUIT BREAKER on the POWER SUPPLY Engage the CIRCUIT BREAKER RTD settings use the SERVICE TOOL Install a new PROCESSOR AY See the ADJUSTMENTS AND REPLACEMENTS THERMAL PROCESSOR AY Extending the PROCESSOR DRUM Install a new DRUM See the ADJUSTMENTS AND REPLACEMENTS DRUM AY THERMAL PROCESSOR AY PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PRO...

Page 230: ...eceive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER ...

Page 231: ...MENTS AND REPLACEMENTS THERMAL PROCESSOR AY Extending the PROCESSOR Check If an existing problem CONNECTORS on the PCB PDB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 232: ...MCS tries to communicate with the MICROCONTROLLER but receives no response Note Error 28 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to communicate with the Processor NORMAL 3 Notify ...

Page 233: ...eive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IM...

Page 234: ... ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem CONNECTORS on the PCB PDB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 235: ...e PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the...

Page 236: ... ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem CONNECTORS on the PCB PDB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 237: ...ing the range limit Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides...

Page 238: ... ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem CONNECTORS on the PCB PDB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 239: ...ER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other ...

Page 240: ... ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem CONNECTORS on the PCB PDB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 241: ...e range limit Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a res...

Page 242: ... ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem CONNECTORS on the PCB PDB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 243: ...to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LA...

Page 244: ... ADJUSTMENTS AND REPLACEMENTS TURNAROUND AY IMAGER with SORTER Check If an existing problem CONNECTORS on the PCB PDB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 245: ...e MCS tries to communicate with the MICROCONTROLLER but receives no response Note Error 29 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to communicate with Densitometer NORMAL 3 Notify ...

Page 246: ...NSITY SENSOR S16 is taken with the light source turned on This occurs when the reading is outside of the optimal range approaching the range limit Note Error 29 925 is generated when this reading is outside the range limit Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Densitometer reference reading outside of optim...

Page 247: ...ENSOR S17 is taken with the light source turned on This occurs when the reading is outside of the optimal range approaching the range limit Note Error 29 927 is generated when this reading is outside the range limit Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm Densitometer fiducial sensor reference reading outside...

Page 248: ... of Error The PRINTER is able to print but films might not go to the correct BIN User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Sorter Internal Communica...

Page 249: ...6 2 Do the following 3 Install a new SORTER See the ADJUSTMENTS AND REPLACEMENTS SORTER AY Check If an existing problem CONNECTORS on the SCB PDB Seat the CONNECTORS SCB Install a new SCB PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 250: ...Result of Error The PRINTER is able to print but films might not go to the correct BIN User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Sorter Internal Com...

Page 251: ...6 2 Do the following 3 Install a new SORTER See the ADJUSTMENTS AND REPLACEMENTS SORTER AY Check If an existing problem CONNECTORS on the SCB PDB Seat the CONNECTORS SCB Install a new SCB PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 252: ...e MCS tries to communicate with the MICROCONTROLLER but receives no response Note Error 30 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No Alarm MCS failed to communicate with Sorter NORMAL 3 Notify ...

Page 253: ... on the IMAGER Result of Error The PRINTER is able to print but films might not go to the correct BIN User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Sort...

Page 254: ...STMENTS AND REPLACEMENTS SORTER AY Check If an existing problem CONNECTORS on the DIVERTER MOTOR SCB PDB Seat the CONNECTORS DIVERTER MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR M16 M19 SORTER AY SCB Install a new SCB PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 255: ...n the IMAGER Result of Error The PRINTER is able to print but films might not go to the correct BIN User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Sorter...

Page 256: ...STMENTS AND REPLACEMENTS SORTER AY Check If an existing problem CONNECTORS on the DIVERTER MOTOR SCB PDB Seat the CONNECTORS DIVERTER MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR M16 M19 SORTER AY SCB Install a new SCB PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 257: ...n the IMAGER Result of Error The PRINTER is able to print but films might not go to the correct BIN User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Sorter...

Page 258: ...STMENTS AND REPLACEMENTS SORTER AY Check If an existing problem CONNECTORS on the DIVERTER MOTOR SCB PDB Seat the CONNECTORS DIVERTER MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR M16 M19 SORTER AY SCB Install a new SCB PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 259: ...n the IMAGER Result of Error The PRINTER is able to print but films might not go to the correct BIN User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Sorter...

Page 260: ...STMENTS AND REPLACEMENTS SORTER AY Check If an existing problem CONNECTORS on the DIVERTER MOTOR SCB PDB Seat the CONNECTORS DIVERTER MOTOR Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR M16 M19 SORTER AY SCB Install a new SCB PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 261: ...ve and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De energize then energize the LASER IMAG...

Page 262: ...g Check If an existing problem CONNECTORS on the PDB and FILM REGISTRATION AY Seat the CONNECTORS FRB Install a new FRB See the ADJUSTMENTS AND REPLACEMENTS FILM REGISTRATION BOARD FRB FILM REGISTRATION AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 263: ...IMAGER Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution...

Page 264: ...Y REGISTRATION CLAMP SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CLAMP S11 FILM REGISTRATION AY CENTERING HOME SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CENTERING HOME S10 FILM REGISTRATION AY REGISTRATION DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTO...

Page 265: ...NTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the othe...

Page 266: ... ACCUMULATOR FILM DETECT S9 FILM REGISTRATION AY REGISTRATION CLAMP SENSOR use the SERVICE TOOL Install a new SENSOR See the ADJUSTMENTS AND REPLACEMENTS SENSOR CLAMP S11 FILM REGISTRATION AY REGISTRATION DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR AY REGISTRATION DRIVE M8 FILM REGISTRATION AY REGISTRATION CLAMP MOTOR use the SERVICE TOOL Install...

Page 267: ...agnostics Result of Error None User Action None FE Action 1 De energize then energize the LASER IMAGER 2 Check and seat all CONNECTORS on the FRB 3 Use the SERVICE TOOL to check the TEMPERATURE RH SENSOR 4 Install a new FRB See the ADJUSTMENTS AND REPLACEMENTS FILM REGISTRATION BOARD FRB FILM REGISTRATION AY LOCAL PANEL LOG DICOM SEVERITY None No alarm Internal Temperature Humidity Sensor Failure ...

Page 268: ... The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Film Registra...

Page 269: ...g Check If an existing problem CONNECTORS on the PDB and FILM REGISTRATION AY Seat the CONNECTORS FRB Install a new FRB See the ADJUSTMENTS AND REPLACEMENTS FILM REGISTRATION BOARD FRB FILM REGISTRATION AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 270: ...e MCS tries to communicate with the MICROCONTROLLER but receives no response Note Error 32 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to communicate with Film Registration NORMAL 3 Notify ...

Page 271: ...INTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Exposure Transport I...

Page 272: ...following Check If an existing problem CONNECTORS on the PDB and ETB Seat the CONNECTORS ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 273: ...rint films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 D...

Page 274: ...ND REPLACEMENTS START OF PAGE SOP BOARD S13 EXPOSURE TRANSPORT AY EXPOSURE ENTRANCE CLAMP MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR ENTRANCE EXIT CLAMP M10 M11 EXPOSURE TRANSPORT AY DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR DRIVE M12 EXPOSURE TRANSPORT AY DRIVE BELT Install a new BELT See the ILLU...

Page 275: ...int films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action Do the actions below in sequence If an action provides a resolution do not continue with the other actions 1 De...

Page 276: ...EMENTS START OF PAGE SOP BOARD S13 EXPOSURE TRANSPORT AY EXPOSURE EXIT CLAMP MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR ENTRANCE EXIT CLAMP M10 M11 EXPOSURE TRANSPORT AY EXPOSURE TRANSPORT DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR DRIVE M12 EXPOSURE TRANSPORT AY DRIVE BELT Install a new BELT See th...

Page 277: ...RINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Exposure Transport ...

Page 278: ...ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem CONNECTORS on the PDB and ETB Seat the CONNECTORS ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 279: ...INTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Exposure Transport I...

Page 280: ...ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem CONNECTORS on the PDB and ETB Seat the CONNECTORS ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 281: ...hen the MCS tries to communicate with the DSP but receives no response Note Error 33 355 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to communicate with Exposure Transport DSP NORMAL 3 Notify ...

Page 282: ...rror The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Exposure ...

Page 283: ...ck If an existing problem CONNECTORS on the DRIVE MOTOR and ETB Seat the CONNECTORS DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR DRIVE M12 EXPOSURE TRANSPORT AY ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION ...

Page 284: ...ror The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Exposure T...

Page 285: ...ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem CONNECTORS on the ETB and PDB Seat the CONNECTORS ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 286: ...of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Expos...

Page 287: ...ck If an existing problem CONNECTORS on the DRIVE MOTOR and ETB Seat the CONNECTORS DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR DRIVE M12 EXPOSURE TRANSPORT AY ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION ...

Page 288: ... The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Exposure Tran...

Page 289: ...ck If an existing problem CONNECTORS on the DRIVE MOTOR and ETB Seat the CONNECTORS DRIVE MOTOR use the SERVICE TOOL Install a new MOTOR See the ADJUSTMENTS AND REPLACEMENTS MOTOR DRIVE M12 EXPOSURE TRANSPORT AY ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION ...

Page 290: ...opped Note Error 33 386 is generated when this problem is detected on 3 successive films Result of Error The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY N...

Page 291: ...The PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Exposure Trans...

Page 292: ...ADJUSTMENTS AND REPLACEMENTS OPTICS AY and EXPOSURE TRANSPORT AY Check If an existing problem CONNECTORS on the PDB and ETB Seat the CONNECTORS ETB Install a new ETB See the ADJUSTMENTS AND REPLACEMENTS EXPOSURE TRANSPORT BOARD ETB EXPOSURE TRANSPORT AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 293: ...e MCS tries to communicate with the MICROCONTROLLER but receives no response Note Error 33 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to communicate with Exposure Transport NORMAL 3 Notify ...

Page 294: ...t new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER 2 Do the following LOCAL PANEL LOG DICOM SEVERITY PDB Internal Communications Failure Alarm sounds PDB microcontroller faile...

Page 295: ...ror is cleared when any DOOR or DRAWER is opened as indicated by INTERLOCKS I1or I3 Result of Error The PRINTER is able to print films but the LATCH must be opened manually when the user needs to access the IMAGER User Action 1 Use the LATCH override mechanism to open the DOOR or DRAWER 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG ...

Page 296: ...JUL07 8F2919 Page 296 of 326 2 Do the following Check If an existing problem CONNECTORS on the LATCH MOTOR LATCH SENSOR Seat the CONNECTORS LATCH SENSOR use the SERVICE TOOL Install a new SENSOR LATCH MOTOR use the SERVICE TOOL Install a new MOTOR ...

Page 297: ...RAWERS could be opened accidentally The error is cleared when a request to unlock is received Result of Error The PRINTER is able to print films but it might be possible for the user to open it without requesting unlock which could cause film exposure or jams User Action 1 Unlock the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize th...

Page 298: ...2919 Page 298 of 326 2 Do the following 3 Install a new Check If an existing problem CONNECTORS on the LATCH MOTOR LATCH SENSOR Seat the CONNECTORS LATCH SENSOR use the SERVICE TOOL Install a new SENSOR LATCH MOTOR use the SERVICE TOOL Install a new MOTOR ...

Page 299: ...obs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER 2 Do the following LOCAL PANEL LOG DICOM SEVERITY PDB Internal Communications Failure Alarm sounds MCS failed to commu...

Page 300: ...e MCS tries to communicate with the MICROCONTROLLER but receives no response Note Error 34 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to communicate with PDB NORMAL 3 Notify ...

Page 301: ...he PRINTER cannot print films The SPOOLER is able to receive and queue jobs but new films will not be initiated Films downstream of the subsystem will be completed All other films will be reprinted User Action 1 Restart the LASER IMAGER from the LOCAL PANEL 2 If the error occurs again call Service FE Action 1 De energize then energize the LASER IMAGER LOCAL PANEL LOG DICOM SEVERITY Processor Trans...

Page 302: ...wing 3 Install a new Check If an existing problem CONNECTORS on the PCB PDB Seat the CONNECTORS PCB Install a new PCB See the ADJUSTMENTS AND REPLACEMENTS PROCESSOR CONTROL BOARD PCB THERMAL PROCESSOR AY PDB Install a new PDB See the ADJUSTMENTS AND REPLACEMENTS POWER DISTRIBUTION BOARD PDB ...

Page 303: ...e MCS tries to communicate with the MICROCONTROLLER but receives no response Note Error 35 931 is generated when this problem is detected after 5 successive communications attempts Result of Error None User Action None FE Action None this is a log notification only LOCAL PANEL LOG DICOM SEVERITY None No alarm MCS failed to communicate with Processor Transport NORMAL 3 Notify ...

Page 304: ...s in area of jam 2 Cycle the transport and check for mechanical problems Failed message displayed on LOCAL PANEL 1 Hardware configuration not correct 2 Internal communications error 1 Use the SERVICE TOOL to check the installed hardware configuration 2 Power cycle the IMAGER IMAGER will not boot up 1 CompactFlash CARD not installed 2 1 Check that the CompactFlash CARD is installed and seated 2 LOC...

Page 305: ...R is open reset A cause might be a short in the POWER MODULE Check POWER SUPPLY outputs 24 V DC and 5 V DC Disconnect the POWER CORD from the wall and use a METER to check the operation of the POWER SWITCH System power is ON then OFF Press the POWER SWITCH on and hold it in If power remains on but de energizes with the SWITCH released check the following 24 V DC POWER SUPPLY CONTACTOR RELAY K1 in ...

Page 306: ...te 1 second ON 1 second OFF MICRO is Happy Operation is normal ON A MICROCODE flash failure occurred during a software update The MICRO is trying recovery OFF A hardware failure has occurred on the MICRO BOARD Blink Code of SORTER HAPPY LED SORTER MICRO State 3 short blinks repeated after a pause Initializing 1 short blink repeated after a pause Idle 2 short blinks repeated after a pause Running 1...

Page 307: ...6 LED SORT_HAZ_V24 present V3_3 Test point for digital 3 3 V V5 Test point for digital 5 V MGND1 MGND2 MGND3 MOTOR ground GND1 GND2 Logic ground V24_HAZ Test point for 24 V hazard DRAWER_HAZARD_24 Test point for DRAWER hazard 24 V SORT_HAZ_24 Test point for SORTER hazard 24 V LASER_HAZARD_24 Test point for LASER hazard 24 V V24 Test point for 24 V S1 Reset SWITCH Item Description The TEST BOX conn...

Page 308: ... point for logic ground S1 Reset SWITCH Item Tests Description The TEST BOX connects to the J1 CONNECTOR on the FSB LED 1 S5 ON film is present at FILM AT FEED SENSOR Note When the film trailing edge passes the SENSOR it goes to 0 V indicating film is present LED 2 ON 5 V present on the BOARD LED 3 S4 ON FEED ROLLER at open position LED 4 S8 ON CARTRIDGE is present LED 5 S7 ON ROLLBACK at open pos...

Page 309: ... for hazard 24 V V5 Test point for digital 5 V V3_3 Test point for digital 3 3 V MGND Test point for MOTOR ground GND1 GND2 Test points for logic ground S1 Reset SWITCH Item Description The TEST BOX connects to the J1 CONNECTOR on the FRB LED 1 ON ACTUATOR_HOME clamped LED 2 ON 5 V is present on the BOARD LED 3 ON CENTER_HOME is at home position LED 5 ON film present at the ACCUMULATOR FILM DETECT...

Page 310: ...hen MOTOR is turning at velocity setpoint D8 LED Programmable LED not used D9 LED Programmable LED flashing indicates DSP is running flashes at 10 Hz rate D10 LED Programmable LED flashing indicates DSP is running flashes at 5 Hz rate D21 LED FILM AT ENTRANCE SENSOR ON film present D25 LED V5 present ON 5 V to BOARD is present D27 LED V24 hazard present ON 24 V present D29 LED SERVO MOTOR 24V ON p...

Page 311: ...ered interval TP19 Test point for ADC5 ITRIP TP20 Test point for ADC0 DAC test TP21 Test point for 5 V DC TP22 Test point for ground TP23 Test point for ENTRANCE MOTOR home TP24 Test point for EXIT MOTOR home TP25 Not used TP28 Test point for ref U digital POT U offset TP29 Test point for ground TP30 Test point for ref V digital POT V offset TP31 Not used TP42 Test point for ground TP43 Test point...

Page 312: ...int for V24 hazard TP51 Test point for ground TP52 Test point for 15 V DC TP53 Test point for 12 V DC TP55 Test point for MOTOR phase U HALL SENSOR TP56 Test point for MOTOR phase V HALL SENSOR TP58 Test point for MOTOR phase W HALL SENSOR S1 Reset SWITCH Item Description ...

Page 313: ...D CENTER EXTRUSION 1 and 2 HEATER zones active ON HEATER active DS7 LED DRUM FRONT HEATER zone active ON HEATER active DS8 LED DRUM REAR HEATER zone active ON HEATER active DS9 LED FLATBED RIGHT EXTRUSION 1 and 2 HEATER zones active ON HEATER active DS10 LED DRUM CENTER HEATER zone active ON HEATER active DS11 LED FLATBED LEFT EXTRUSION 3 HEATER zone active ON HEATER active DS12 LED DRUM zone retu...

Page 314: ...CH Item Description D1 LED Happy LED V5 Test point for digital 5 V MGND Test point for MOTOR ground GND Test point for logic ground SORT_HAZ_V24 Test point for hazard 24 V Item Description The TEST BOX connects to the J2 CONNECTOR on the SCB LED 1 ON DIVERTER 4 at home position LED 2 ON 5 V is present on the BOARD LED 3 ON DIVERTER 3 at home position LED 5 ON DIVERTER 2 at home position LED 7 ON D...

Page 315: ...rrors from installation of the system IIS INTERNET INFORMATION SERVER Log This log records all attempts to connect to the Web site SERVER LOCAL PANEL Log This is a log of events for the LOCAL PANEL SUBSYSTEM filtered from the Application Log The LOCAL PANEL SUBSYSTEM is part of the MIM Application MCS Log This is a subsystem log of MCS events filtered from the APPLICATION Log The log records all i...

Page 316: ...NK CLIENT SOFTWARE to access the SERVICE TOOL Doing a search of the ERROR TALLY to determine the number of times the error has occurred Opening the NT EVENT System Log to determine the time at which the error occurred Doing a search of the APPLICATION Log or a subsystem log for data at the time of the error Viewing the data in the log before the error If necessary saving the log and sending it to ...

Page 317: ...hat are of no use The items checked above are the most used for troubleshooting in the field The other items are for Software Engineering use Note You can also decrease or increase the column width in the log by selecting the lines between the columns and moving the column left or right 6 Click OK Log Keying Functions 1 Start Click Start at any time to display parameters for retrieving a new log T...

Page 318: ...ea in the log that you can identify by words or numbers for example an error code Enter the combination of words and numbers you want to locate in the Find what window Click Find to search the log 5 Save Click Save to save a log to a folder on your LAPTOP Use the cursor to select the area of the log you want to save Select the folder on the LAPTOP for example Log Files Enter a File name and click ...

Page 319: ...r Test contains 5 Type the words and numbers that you want to search on In this example scp assoc was selected to appear in the events in the log you want to view 6 Click OK to search the log for the selected test 7 Note that the search retrieved only log items that include the words scp assoc Also note that the log includes 1 688 records To decrease the data do additional filtering 8 Click Displa...

Page 320: ... be removed from the viewed log by the Options function described in the preceding section The columns from left to right include 1 Event Type This column provides the same information in the Level column It describes the level provided for the event The levels from highest to lowest are TRACE Level 6 DEBUG 5 DIAG 4 INFO 3 WARNING 2 ERROR 1 and FATAL ERROR 0 2 Time Provides the time of the event d...

Page 321: ...stem negotiates with the SCU for sending a job from the SCU across the network It then makes a Film Session sets up the Film Box and Images Box These will appear in the first pages of the APPLICATION Log IMPORT This subsystem receives the images from the network and processes the images for storage on the HARD DRIVE STORAGE This subsystem is for temporary storage of images on the HARD DRIVE DELIVE...

Page 322: ... 0 Fatal Error Record only fatal errors 1 Error Record all errors 2 Warning Record warnings and all errors 3 Information Record key events 4 Diagnostic Provide a level of information for field troubleshooting 5 Debug Provide a level of information for software debugging 6 Trace Important Levels higher than the recommended defaults can slow down the system If you change a level to help diagnose a p...

Page 323: ...ve occurred during the operation of the IMAGER 1 To access these logs connect your LAPTOP to the SERVICE TOOL and select Logs Error Tally The Current Error Tally will display Note The error codes in the tally are by application DV6800 MIM or Service Platform a To reset the count for any error to 0 click Select for that error and then click Reset at the bottom of the screen b If you want to reset t...

Page 324: ...L Log This is a subsystem log that includes all events for the LOCAL PANEL filtered from an Application Log See Sample Log 1 at the end of this section for descriptions of the events in the log MCS Log This is a subsystem log that includes all MCS events pulled from an Application Log These are events of any of the MICRO subsystems on the MCS I2 C bus Note that each description of the event begins...

Page 325: ...vents for system software PRINT DELIVERY Log This is a subsystem log that includes all DELIVERY and SDS SERVER SUBSYSTEM events pulled from an Application Log Note that the subsystem column includes only these subsystems See Sample Log 1 at the end of this section for descriptions of the events in the log ...

Page 326: ... Carestream Health Inc 150 Verona Street Rochester NY 14608 Section 5 Publication History Publication Date Publication No ECO No Changed Pages File Name Notes 30JUL07 8F2919 CN0008637 8f2919 fm New Publication Rev A DryView is a trademark of Carestream Health Kodak is a trademark of Kodak used under license ...

Reviews: