DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
114 of 326
Error 20-551: PROCESSOR FILTER Maintenance Necessary
Messages
Description
The PM message is written to the log when the sheet count exceeds the sheet count limit for
the Preventive Maintenance Module (PMM). An indicator on the LOCAL PANEL is used to
notify the operator of the PM requirement.
Result of Error
None
User Action
1
If at 15,000, 45,000, 75,000, etc. Replace PMM MODULE.
FE Action
1
Change the PMM.
LOCAL PANEL
LOG
DICOM
SEVERITY
None
(No Alarm)
Processor Filter Maintenance
Required
NORMAL
3-Notify