DIAGNOSTICS
Using the Diagnostics
30JUL07
8F2919
Page
14 of 326
•
stop the agent, click [Stop]
•
start the agent, click [Start]
•
restart the agent, click [Restart]
7
Click [Save].
Diagnostic Tests
DENSITOMETER
1
Select
Diagnostics>MCS>Densitometer
.
2
Click [Run].
Note
This is the same test run at power ON for the DENSITOMETER. The test ends with a “Pass”
or “Fail” indication and displays diagnostic data. The test sequence:
•
De-energizes the DENSITY and FIDUCIAL LAMPS, then reads A to D counts for offset
light values for the LAMPS 8 times and calculates an average. The average value must be
4096 + 400.
•
De-energizes the DENSITY LAMP, then reads the A to D counts for reference brightness
8 times and averages the values. If the average A to D count is not between 750,000 and
850,000, software adjusts the count.
•
De-energizes the DENSITY LAMP and energizes the FIDUCIAL LAMP. Reads the
Reference brightness value 8 times and averages the values. If the average value is not
between 250,000 and 550,000, software adjusts the brightness.
3
To display the results of the last DENSITOMETER test, click [Current Diagnostic Data].
4
To save the test results to a file on your LAPTOP COMPUTER, click [Export Data].