DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
190 of 326
Error 27-602: Invalid POWER MONITOR Range with ATTENUATOR Open
Messages
Description
This error occurs when the POWER MONITOR reading with the ATTENUATOR wide open
measured by the OPTICS calibration and diagnostics is outside the range. The error clears
when a calibration is asked for or power is cycled.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
DICOM
SEVERITY
Optics: Calibration Failed
(Alarm sounds)
Optics Calibration Failed.
Power Monitor value with
Attenuator Open outside valid
range
ELEC DOWN
0-Fatal