DIAGNOSTICS
Status Messages and Error Codes
30JUL07
8F2919
Page
198 of 326
Error 27-606: ATTENUATOR Calibration Failed (Non-Monotonic)
Messages
Description
This error occurs when the BEAM POWER measurements taken during ATTENUATOR
calibration are non-monotonic. The error clears when a calibration is asked for or power is
cycled.
Result of Error
The PRINTER cannot print films. The SPOOLER is able to receive and queue jobs but new
films will not be initiated. Films downstream of the subsystem will be completed. All other
films will be reprinted.
User Action
1
Restart the LASER IMAGER from the LOCAL PANEL.
2
If the error occurs again, call Service.
FE Action
Do the actions below in sequence. If an action provides a resolution, do not continue with the
other actions.
1
De-energize then energize the LASER IMAGER.
LOCAL PANEL
LOG
DICOM
SEVERITY
Optics: Calibration Failed
(Alarm sounds)
Optics Calibration Failed.
Power Monitor readings not
monotonic
ELEC DOWN
0-Fatal