4-8
5E2234
June 2000
Calibration
13.
Press the
Menu
Button at the density value to be
changed. The cursor will flash, which will allow you to
change the values. Use the “
▼
”
or “
▲
“ to change the
value. Press the
Menu
Button to accept the change.
14.
Use the “
▼
”
to move the “
>”
to
Exit
. Press the
Menu
Button.
15.
Repeat steps 12 - 14 for all densities to be changed.
16.
Use the “
▲
“ to move the “
>
”
to
Exit
. Press the
Menu
Button to return to the Calibration Densities Screen.
17.
Use the “
▼
”
to select “Compute Calibration”.
18.
Press the
Menu
Button to complete the calibration.
The calibration information is erased from the screen
display, and is replaced by “Computing Film Model”. If
the calibration is not successful, see Chapter 5
“Troubleshooting” for more information.
19.
Use the “
▲
“
to move the “
>
”
to
Exit
. Press the
Menu
Button 2 times to return to the Main Menu.
Ready
Density Step 1
>0.14 _
Exit
Cancel Change
Menu to change
Ready
Receive Magazine
Last Calibrate on
13-Dec-1998 07:24
>Exit
Density Values
Compute Calibration
Ready
Receive Magazine
Calibration Complete
13-Dec-1998 07:24
>Exit
Density Values
Compute Calibration