User’s Manual
6–8
5B9615
[5] Press [F4] to read the Calibration Test Film. The Printer will:
• Clear all commands and change the prompt to “Please wait.”
• Change the prompt to “Slide calibration film through the densitometer.”
You must slide the calibration film through the Densitometer within about 15 seconds.
Use an even sweeping motion when sliding the film, otherwise a calibration error may
occur.
When you slide the film through the Densitometer, the Printer displays the message,
“Calibration in progress; please wait.”
If the calibration is successful, the Printer changes the prompt to “Printer calibration
successful.”
If any errors occur during calibration, the Printer will display the Error Recovery
Screen. The existing calibration table is not affected. Printed Errors and the Error
Recovery Screen will be discussed in detail in Section 8, “Troubleshooting and Error
Messages.”
Press [F6] to return to the previous screen.
[6] Measure the density values at an off-line Densitometer and then press [F5] to enter
the density data at the Control Panel.
1
2
3
4
5
6
7
8
9
H126_0022AA
test film
calibration
densitometer
H126_0022ACB