User’s Manual
8–18
5B9615
You must slide the Reference Strip through the Densitometer within about 15 seconds.
Use an even, sweeping motion when sliding through the Densitometer.
When you slide the Reference Strip through the Densitometer, the Printer displays the
message, “Densitometer calibration successful.” The Printer will return to the Display
Screen as shown above.
If any errors occur during calibration, the Printer displays the Error Recovery Screen.
H126_0023AA
strip
reference
densitometer
H126_0023ACA