Processor Overview
5B9615
11–7
Processor Configuration and Setpoints
Setup information consists of the film processor configuration and process setpoints.
Setup information that was preset at the factory is listed below.
Setup information is stored in battery backed-up RAM (Random Access Memory).
Therefore, you do not need to program new values every time that you turn on the
Processor. Even if power to the Processor is interrupted or the Processor is turned
off, you do not need to program the Processor.
If you wish, you may change the setup information by using the function keys on the
display panel. See Section 13, “Setting Up the Processor,” for more information. The
new setup information will be retained in memory when power is interrupted or turned
off.
The Processor configuration and setpoints are initially set up as follows:
Replenishment Mode
Automatic
Display Units
°
F
Developer Temperature
37.8
°
C
(100
°
F)
Fixer Temperature (minimum)
37.8
°
C
(100
°
F)
Automatic Developer Replenishment
(35 x 43 cm sheet)
60 mL
Automatic Fixer Replenishment
(35 x 43 cm sheet)
85 mL
Flooded Developer Replenishment
(35 x 43 cm sheet)
65 mL
Flooded Fixer Replenishment
(35 x 43 cm sheet)
65 mL
Dryer Temperature
48.9
°
C
(120
°
F)
Summary of Contents for Ektascan 2180
Page 70: ...User s Manual 3 44 5B9615 ...
Page 100: ...User s Manual 4 30 5B9615 ...
Page 116: ...User s Manual 5 16 5B9615 ...
Page 162: ...User s Manual 8 20 5B9615 ...
Page 164: ...User s Manual 9 2 5B9615 ...
Page 178: ...User s Manual 12 2 5B9615 ...
Page 202: ...User s Manual 14 6 5B9615 ...
Page 226: ...User s Manual 16 12 5B9615 ...
Page 258: ...User s Manual 18 20 5B9615 ...