Flooded Replenishment Mode
In the Flooded Replenishment Mode, the preset values for developer and fixer
replenishment volumes are:
• 65 mL every 5 minutes.
• 65 mL for each 1500 cm
2
(238 in.
2
) of film.
If you wish, you may change the setup information by using the function keys on the
display panel. See “Changing Setpoints” in this section.
Important
Before changing the replenishment volumes, calibrate the replenishment system.
See “Calibrating the Replenishment System” in this section.
Press [F2] at the Processor Setup Screen to display the Processor Replenishment Mode
Screen.
PROCESSOR REPLENISHMENT MODE
Processor Ready
F2
F3
F4
Return
F6
Select the desired function.
Disable Replenishment
Fill Tanks
Select Flooded
Replenishment Mode is Auto
F1
H126_9053AC