User’s Manual
13–10
5B9615
Calibrating the Replenishment System
Procedure
Important
Calibrate the replenisher pumps every 3 months.
Calibration of the replenishment system will determine the actual rate of solution flow
through the Replenishment Pumps for a given period of time.
Measure the volume of solution that is delivered by the system, then use the Display
Panel to enter the measurement into the microprocessor. The microprocessor computes
the rate of solution flow through the Pump, then adjusts the period of time that the Pump
must operate to deliver the replenishment volume that was set.
Note
The volume actually measured during this procedure is not the volume that is delivered
for a 35 x 43 cm sheet of film during normal equipment operation.
[1] To calibrate the replenishment system, press [F4] at the Other Functions Screen to
see the Control Processor Screen.
[2] Press [F3] at the Control Processor Screen to display the Processor Setup Access
Screen.
[3] Enter the four-digit Access Code: 4213.
PROCESSOR SETUP
Processor Ready
F2
F3
F4
F5
Return
F6
Select the desired action.
Fix Replen Calibration
Dev Replen Calibration
Setup Process
Change Replen Mode
Cycle is Rapid
Replenishment Mode is Auto
Processor Error: E134
F1
H126_9055AC
Summary of Contents for Ektascan 2180
Page 70: ...User s Manual 3 44 5B9615 ...
Page 100: ...User s Manual 4 30 5B9615 ...
Page 116: ...User s Manual 5 16 5B9615 ...
Page 162: ...User s Manual 8 20 5B9615 ...
Page 164: ...User s Manual 9 2 5B9615 ...
Page 178: ...User s Manual 12 2 5B9615 ...
Page 202: ...User s Manual 14 6 5B9615 ...
Page 226: ...User s Manual 16 12 5B9615 ...
Page 258: ...User s Manual 18 20 5B9615 ...