User’s Manual
13–18
5B9615
Standby Mode
If the 180 LP or 180 LPS Processor doesn’t receive film for approximately 60 seconds
after the last film has exited, the Processor will enter standby mode. When the Processor
goes into standby mode, the following events occur:
• the water supply is turned off, unless it is needed to cool the developer.
• the developer is maintained at setpoint temperature. The fixer is maintained at
setpoint temperature or higher, since it is not cooled.
• the Dryer Heater and Blower turn on and off at a fixed interval of 45 seconds. The
Blower is turned on and the temperature is checked. If necessary, the Heater turns
on until the setpoint temperature is reached. Once the proper temperature is reached,
the Blower and Heater turn off and stay off for 45 seconds when the cycle is
repeated.
• the Drive Motor and water supply will turn on and off every 8 minutes for 1
1
⁄
2
minutes to keep the Rollers wet.
The Processor will come out of standby mode and operate when a film is printed and sent
to the Processor.
Display Units
When your system is installed and configured, the installer can select the temperature
display format for English (
°
F), or metric (
°
C) temperature units. If you need to change
the display units, contact your Service Representative.
Summary of Contents for Ektascan 2180
Page 70: ...User s Manual 3 44 5B9615 ...
Page 100: ...User s Manual 4 30 5B9615 ...
Page 116: ...User s Manual 5 16 5B9615 ...
Page 162: ...User s Manual 8 20 5B9615 ...
Page 164: ...User s Manual 9 2 5B9615 ...
Page 178: ...User s Manual 12 2 5B9615 ...
Page 202: ...User s Manual 14 6 5B9615 ...
Page 226: ...User s Manual 16 12 5B9615 ...
Page 258: ...User s Manual 18 20 5B9615 ...