User’s Manual
16–4
5B9615
Entering Density Data into the Kodak Process Control Densitometer
[1] Use [
↑
] and [
↓
] to select the channel for storing the processor monitoring values.
[2] Orient the 21-step gray scale pattern film, with its emulsion side down, for
insertion into the Densitometer as illustrated.
[3] Position the film against the Stop of the Densitometer. Insert the film until it rests
against the Drive Rollers and the Motor is activated.
[4] While the film density is measured, gently hold the film against the Stop, guiding
the film to prevent skewing.
[5] Press [go]. The date and time are momentarily displayed.
[6] Use [
↑
] and [
↓
] to enter the temperature of the developer solution.
[7] Press [go] to store the data and advance to the next step.
[8] Press [meas.] to view the calculated parameters and density values measured on the
21-step gray scale, or press [dev.] to view the data as deviated from the reference
values.
[9] Press [p1] through [p4] to view the measured data.
[10] Press both “Menu” keys to exit the main menu.
Note
If the Densitometer is connected to an appropriate Printer, the data can be sent to the
Printer. Refer to Section 3.3 of the Operation Manual for the Kodak Process Control
Densitometer, Publication Part Number 635933, for further information on sending data.
Summary of Contents for Ektascan 2180
Page 70: ...User s Manual 3 44 5B9615 ...
Page 100: ...User s Manual 4 30 5B9615 ...
Page 116: ...User s Manual 5 16 5B9615 ...
Page 162: ...User s Manual 8 20 5B9615 ...
Page 164: ...User s Manual 9 2 5B9615 ...
Page 178: ...User s Manual 12 2 5B9615 ...
Page 202: ...User s Manual 14 6 5B9615 ...
Page 226: ...User s Manual 16 12 5B9615 ...
Page 258: ...User s Manual 18 20 5B9615 ...