A-61096 June 1999
5-3
3. Locate and grasp the end of the lower imaging guide. Carefully
slide the imaging guide out of its track.
4. Remove any dust from the imaging guides using a soft bristle
brush.
5. Remove fingerprints or smudges from the imaging guides, wiping
both sides of each imaging guide with a clean, soft, lint-free cloth
slightly moistened with water or lens cleaner. Lightly wipe both
sides of each imaging guide with a dry cloth or an antistatic wipe.
IMPORTANT:
Do not use cleaning pads on the imaging guide s.
6. Remove the upper lamp housing assembly.
7. Remove any dust from the upper lamp housing assembly using a
soft bristle brush.
8. Reinstall the upper lamp housing assembly.
9. Repeat Steps 6 through 8 for the lower lamp housing assembly.
10. Grasp and hold the lower imaging guide by the green handle and
carefully slide it back into place. Make sure the grooves fully
engage the track.
11. Grasp and hold the upper imaging guide by the green handle and
carefully slide it back into place. Make sure the grooves fully
engage the track.
NOTE: The imaging guides are not interchangeable. They can
only be inserted one way. Use the illustration on the
previous page to assist you in properly positioning the
imaging guides.
12. Close the right-side access door.