Diagnostics and Troubleshooting
September 2000
5-23
601
A bad count was entered.
The expected count
was:_count_.
• A file was edited manually and
an incorrect number of rows
was entered.
• An internal software error
occurred.
• Manually enter the correct
number of rows in the file.
• If problem continues, call for
service.
602
An error occurred during the
calculation process.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
701
Patch _patch_, column
_column_, value _value_ is
beyond tolerance
_tolerance_ for aim _aim_.
A density patch was out of
tolerance.
Perform another calibration cycle.
702
Patch _patch_ has a spread
of _spread_ which is out of
range _range_.
The spread between colors for a
given density patch is out of
tolerance.
Perform another calibration cycle.
703
CalToleranceFile can only be
initialized once.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
704
CalToleranceFile::
initialize must be called first.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
801
Could not find row:_row_.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
802
Could not find
column:_column_.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
803
Could not find element -
column:_column_.
row:_row_.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
804
Row value was not set
before using.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
805
Column value was not set
before using.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
806
Error deleting row:_row_.
An internal error occurred with the
software.
• Restart the application.
• If problem continues, call for
service.
807
Element at row:_row_;
column:_col_
value:_val_ failed audit.
An input file had incorrectly
formatted data in it at the location
indicated.
• Verify that you are using the
correct file.
• Edit the file and correct the
format.
Error
Code
Error Code /Status
Message
Possible Cause/Subsystem
Possible Solution
Summary of Contents for PROFESSIONAL LED II 20P
Page 16: ......
Page 20: ......
Page 74: ......
Page 90: ......
Page 118: ......
Page 122: ......
Page 146: ...Additional Calibration Information C 12 September 2000 Value Graph ...
Page 147: ...Additional Calibration Information September 2000 C 13 Difference Graph ...
Page 168: ......
Page 190: ......
Page 197: ......