System Operating Procedures
3-36
September 2000
Viewing/Setting Replenishment Settings
Viewing the Replenishment Rate
1.
Press Replenishment to access the Replenishment screen.
2. Press
the
Up or Down Arrow to select the desired replenishment pump or the
wash/stabilizer tanks.
Setting the Chemical Replenishment Rate
1.
Select the desired replenishment pump.
2. Press
Set to access the Set mode.
3. Use
the
Up or Down Arrow to change the replenishment needed for each
square foot of paper processed.
4. Press
Set twice or press a different blue function key to return to View mode.
NOTE: Press Replenishment to toggle the pump to Manual on, Manual off or
Per-Square-Foot mode.
Setting the Wash Replenishment Rate
1.
Select the desired wash/stabilizer tank.
2. Press
Set to access the Set mode.
3. If the wash is in per-square-foot mode, use the Up or Down Arrow to change
the amount of wash solution needed for each square foot of paper processed.
NOTE: In Automatic mode, the wash solution is on whenever material is in the
processor.
4. Press a different blue function key to return to View mode.
NOTE: Press Replenishment to toggle the pump to Manual on, Manual off,
Per-Square-Foot, or Automatic mode.
Calibrating the Replenishment Flow Rate
NOTE: Turn the key switch to the service position before you calibrate the
replenishment flow rate.
IMPORTANT: Before calibrating a flow rate, position the desired replenishment
tube into a beaker to catch the pump’s output or the wash tank’s
overflow. Make sure the beaker is large enough to catch the output
of the pump for at least 10 replenishment bursts of one second
duration. The pump may make more than one revolution during a
single burst period.
Measuring the Flow Rate
1.
Select the desired replenishment pump.
2. Press
Set twice to access the Replenishment screen in the Calibrate mode.
3. Press
Replenishment to initiate a 10-second auto calibrate cycle for the
selected pump. After the cycle has completed, measure the output and divide
it by ten. This is the actual flow rate.
Summary of Contents for PROFESSIONAL LED II 20P
Page 16: ......
Page 20: ......
Page 74: ......
Page 90: ......
Page 118: ......
Page 122: ......
Page 146: ...Additional Calibration Information C 12 September 2000 Value Graph ...
Page 147: ...Additional Calibration Information September 2000 C 13 Difference Graph ...
Page 168: ......
Page 190: ......
Page 197: ......