System Operating Procedures
September 2000
3-41
Setting the Development Time Tolerance
A “DRIVE ERROR” message appears on the OCP when the development time is
slower or faster than the set speed by more than its set tolerance. The information
in this procedure enables you to change the tolerance.
1.
Use the Up or Down Arrow to select the desired setup (A or B) and the
tolerance screen.
2. Press
Set to access the screen in Set mode.
3. Use
the
Up or Down Arrow to adjust the acceptable tolerance range. The
adjustments are in one second increments.
4. Press
Set to cancel the Set mode.
5. To save any changes and return to normal operation, press both Reset keys
simultaneously.
NOTE: To make other changes, select the desired screen and enter new
values.
Setting the Chemical Temperature Tolerance
This must be done by a qualified Kodak service person.
Clearing Error Conditions
Refer to the following procedures to clear error conditions:
•
“Clearing Drive Errors” on page 5-18
•
“Clearing Processor Paper Jams” on page 5-18
•
“Clearing a Reset Message” on page 5-17