12.Current
consumption
I
stand-by
<2μA; I
working
<0.5mA; I
Buzzer on
<2mA
;
I
banklight
<15mA
13.LED backlit
specifications
Red 1.2cd/m
≧
2
Green
≧
2.0cd/m
2
15.Operational
conditions
ASTM 10 -40 (50 -104 )/15-95%RH
℃
℃
℉
℉
16.Type of
measuing
Applicable for forehead temperature and milk
temperature
17.battery
Changeable for two 1.5V triple A batteries
18.Battery life More than 1000 times
19.Accuracy for
clinical test
The maximum allowable error for clinical
test is specified in the formula below:
0.3 /0.6 (for 95%)
≦ ℃
℉
Among which:
T1 and T2 represent temperature value for
thermometer under test respectively, Tref
represents the constant reference temperature
※
It is recommended that any professional users apply technical
inspection once every two years in accordance with Medical
Products User Law.
Error
=
T1-Tref
T2-Tref
+
2
Summary of Contents for KATMRIRNONA
Page 1: ......