26
Stimulation is
uncomfortable
1. Intensity is too high
2. Electrodes are too close
together
3. Damaged or worn
electrodes or lead wires
4. Electrode active area size
is too small.
1. Decrease intensity.
2. Reposition the electrodes.
3. Replace.
4. Replace electrodes with ones that have an
active area no less than 16.0cm
2
(4cm*4cm).
Intermittent output
Lead wires
Program option in use
1. Verify connection is secure. Insure firmly.
2. Turn down the intensity. Rotate lead wires in
socket 90°. If still intermittent, replace lead
wire.
3. If still intermittent after replacing lead wire, a
component may have failed. Call the
Kogan
customer support team
.
Some programs will seem intermittent. This is
expected.
Refer to the Program Option Controls in the
Operation section for a description of the
program option.
Stimulation invalid
Electrodes and the application
position is abnormal
Unknown
Change electrodes and the application position
Contact your doctor
9. Storage
1. Remove the batteries from the unit if you are not going to use it for a longer period. Leaking
batteries can damage the unit.
2. Store the device in a dry room and protect it against heat, sunshine and moisture.
3. Store the device in a cool, well-ventilated place
4. Detach the connecting cable from the electrodes.
5. Never place any heavy objects on the device.