AVV-711 Quick Book
75
13.4 LIVE MOVE
For live adjustment of any selected plane place detector at the center of that bore the fixture should be in
fixed position – at the bottom side of the bore(180º).
In result view screen select reference line mode and switch to table view mode (if not), select desired
plane using arrow keys
, then press
Measure screen with live move ready state appears
If inclinometer does not used you should manually enter angle first.
Press
again to start live move.
Do not move any parts until blinking “LIVE MOVE” showed!
Move object toward zero result. To stop live and take result after moving press
Do not stop live
while measure data not ready (beam out of target or hourglass or connection error)
After live move stopped change plane ability become available
To live move other plane change plane number to the new selected
plane, place detector at the center of the new desired bore and
press
to ready to live move, then press
to start live move
Adjustment of one plane can affect other planes, so repeat
measurements after live move (go to
result
screen then press
to
remeasure)
Live ready but
not started
Live move enabled.
adjust now
live result,
move toward
zeroes
[*] means plane live
moved