AVV-711 Quick Book
107
16.3 VIEW AND SAVE RESULT
When measurement complete for all positions press
for result.
There are two reference line modes. Press
to switch between them.
Ref.points
mode: when any of the reference points undefined result is the unchanged detector value
(absolute plumbline)
when two reference points are both properly defined result is the difference between calculated reference
line and the detector value;
To define reference points press
then enter reference point 1 number and reference point 2 number
respectively. Press
for store number. To clear reference point enter decimal point
as value
Best fit
mode : In the best fit mode result is the difference between calculated best fit reference line and
the detector value.
The result can be viewed as the table or as graph. Press
to change view.
For save result to file press
. See
For return to main menu press
Cautions in this screen
Shortcuts in this screen
Tasks in this screen
To switch view between table/graph
·
Taking
measurements
·
Manually enter data(if manual data
chosen in parameters)
To enter reference points(use
for
confirm input)
To switch reference line mode
To save to file
To repeat measurement(all data will be
lost)
To go to measure