Manual EDU 2AE/TOP/MITO
Page 9
1) Language:
This option allows to choose among 6 languages: English, Italian, German, Spanish, French
and Portuguese.
2) Change password:
You can change password (see point 3).
- - - - O P T I O N - - M E N U - - - -
3 ) P a s s w o r d :
O N
4 ) E S C - O K :
I N T + E X T
5 ) R e s e t :
P R G
3) Pass ON:
If YES, you will be asked for the password to enter the menu. Otherwise select NO.
ATTENTION: Password is only asked to enter the menu the first time after switching the unit on.
The default password is: 0000.
4) ESC-OK:
Press ESC in case of error. Press OK at cycle end or to reset a cycle or a program (see point
5: Reset). If INT,
the operator must press the required key on the front panel. If it’s set on EXT, the keys on
the front panel are disabled and functions can only be enabled through proper connections on the back panel
(
see paragraph “I/O connections”).
5) Reset:
you can select which parameter resets when you press RESET (see point 4: Keyboard
commands). If PRG, the Reset refers to the program that is being carried out. If SEQ, the Reset refers to the
whole sequence (only if sequence is enabled).
- - - - O P T I O N S - M E N U - - - -
5 ) R e s e t :
P RG
6 ) S e t
t i m e
&
d a t e
7 ) C y c l e s :
0 0 0 0 0 0 0
6) Set time & date:
to set time and date on the control unit.
7) Cycles
- Total number of cycles: It shows the number of tightened and loosened screws. This number
can’t be modified
.
TORQUE CALIBRATION
Please read the instructions carefully and equip a torque tester to proceed.
The torque calibration function allows to view the torque value in Nm or in.lbs directly on the display.
ATTENTION: The
displayed torque level is approximate and it’s determined by mathematical calculations.
The smaller the calibration interval the more precise the value on the screen.
IMPORTANT WARNINGS:
Calibration values will be referred to the parameters of the control unit which are set during
calibration. If you modify speed or type of joint, the unit will need to be recalibrated.
If the screwing cycle during calibration isn’t carried out correctly or if you’re not sure of the value
it’s possible to repeat it. A wron
g screwing cycle can affect the precision of the calibration cycle.
Calibration is carried out for two torque levels optionally chosen. The minimum calibration interval
is 5% (example: if I have to work at 30%, I can calibrate at 25% and 35%) but we recommend a
minimum of 10% interval.