Ch.3 Setup
< 101 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
1
Switch to the service maintenance mode.
p.86 Switching to
the Service
Maintenance
Mode
2
Touch the [SERVICE 1] button.
- The installation setup buttons will be displayed.
3
Touch the [SCU SETUP] button.
- An input device selection screen will be displayed.
- The following three input devices are set as default.
CH01
: Regius 170
CH02
: Printlink
CH03
: Regius 190
To change information for these channels, touch
the corresponding device button.
After inputting new input device information, touch
the button corresponding to the channel to which
the information is to be added.
5
Set input device usage, input device name, IP
address and host name.
- To make input device usage settings, touch the
"ON" or "OFF" button.
- Touching the input device name, IP address or host
name buttons will produce display of a
corresponding set value input screen.
p.84
Set Value
Input Screen
6
Touch the [>] button at the bottom left of the
screen.
- The SCU setting screen (2/4) will be displayed.
4
Touch the channel button to be set.
- The SCU setting screen (1/4) will be displayed.
2
3
4
5
6