Ch.3 Setup
< 102 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
7
Set the input device N-EVENT port number, AE
title and DICOM error level.
- Touching the input device N-EVENT port number,
AE title buttons will produce display of a set value
input screen.
- Use the [<][>] buttons to switch between DICOM
error level settings.
8
Touch the [>] button at the bottom left of the
screen.
- The SCU setting screen (3/4) will be displayed.
9
Make size/type check, forced specified film use
and used film size/type settings.
- Use the ON/OFF buttons to enable or disable film
size/type check and forced specified film use
settings.
- Use the [<][>] buttons to switch between screens
for used film size/type settings.
10
Touch the [>] button at the bottom left of the
screen.
- The SCU setting screen (4/4) will be displayed.
11
Set the empty tray notification.
- Use the ON/OFF buttons to enable or disable empty
tray notification settings.
12
Touch the [SAVE] button.
- Settings will be saved and display returned to the
service maintenance screen.
7
8
9
10
11
12