Ch.3 Setup
< 123 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Touching the [Return] button during the
process of making setting changes will
produce display of a message requesting
confirmation of whether or not to save
settings.
[Yes]
: Settings will be saved and display
returned to the service
maintenance screen.
[No]
: The service maintenance screen
will be displayed without saving
settings.
1
Select the LUT number to registered by touching
the scroll buttons at the left and right of the edit
LUT number.
(2) Registration of LUT data read from the DRYPRO 793
internal library after making fine adjustments.
The procedure for registration of LUT data read from the DRYPRO 793
internal library after making fine adjustments is described below.
3
Using the left/right scroll buttons, select the
diagnostic device.
- The LUT name corresponding to the selected
device will be displayed.
4
Touch the LUT name to be registered.
2
Touch the [REFERENCE LIBRARY] button.
- The LUT library screen will be displayed.
1
2
3
4